22
Version 4.2 rev 06 Oct 2020
8.2 x 8.2 mm / 512 x 512 sensor
Lowest Noise Imaging EMCCD
Mechelle
2.3 e
ntRance
a
peRtuRe
s
lit
The standard Mechelle is supplied with the SMA connector and brass tool (
Figure 3, left)
. The maximum aperture height
recommended to avoid order crosstalk is 50 μm. The entrance aperture can be either, a direct fibre input, a slit or a
pinhole. When a fibre-optic input is used, the best arrangement for maximum throughput is to have the fibre itself as the
aperture. In that case the fibre diameter should not exceed 50 μm.
In some applications fibre-optics cannot be used, particularly in the UV region, due to the lack of throughput, so a
Slit
Disc
must be used (
Figure 3, right
).
Figure 3: (Left) SMA connector and brass insertion tool, (Right) Slit disc and brass insertion tool
The most common slit size used is a rectangular 25 μm x 50 μm (W x H) slit. Other apertures such as a 50 μm x 50 μm, or
50 μm pinhole can also be used. However, these wider slits will provide a lower resolution than the 25μm x 50μm slit, but
yield a higher throughput. The maximum recommended slit width is 200 μm.
NOTE: If a 10 μm x 50 μm slit is used the resolution of the system will improve but throughput will be reduced.
The exchange procedure for the SMA and other adapters is shown in
Section 6.3
.
2.3.1 O
ptiOnal
St & FC
Fibre
adapterS
Optional
ST and FC
fibre adapters can also be used with the Mechelle spectrograph.
NOTE: After changing slit to fibre, the unit must be recalibrated.
Figure 4: (left) FC Connector, (right) ST Connector