background image

Connecting your Camera and Accessories to the HoloSpec

4

Refer also to your camera user manual

Setting up Solis for the HoloSpec

5

1.  Insert the power cable from the power supply into the power 

connector at the rear of the camera. Secure in position using 

the screwlock.

2.  Insert the mains cable into the camera power supply unit.
3.  Connect the USB cable to the camera and an available USB 

port on the PC.

4.  If a shutter is installed, connect the SMB to BNC trigger cable 

between the camera and the shutter driver.

5.  Switch on the camera.
6.  Start your software.
7.  View and adjust the acquisition settings as required.
8.  Start acquiring data.

Troubleshooting

Camera is not recognized by PC

•  Check that the camera power and USB cables are connected securely and the mains power cable is inserted.

•  Check that the drivers have been installed properly 

(Control Panel /System/Hardware /Device Manager /Imaging Devices).

Connecting your Camera and Switching it on

Ensure that the Camera ON/OFF switch is in the OFF 
position AND your PC is switched OFF.

6

MAN-08932

HOSG 1217 R1

For additional troubleshooting information please refer to the user manual
For further technical information about the Andor HoloSpec go to: 

www.andor.com/spectrographs

You can find the latest user manuals on 

MyAndor (http://my.andor.com/login.aspx)

?

Refer also to your Solis software guide

1.  Launch your Solis Software.
2.  Navigate to 

Hardware/Spectrograph Setup

.

3.  Select “Andor” in the 

Manufacturer

 field.

4.  Select “HoloSpec” in the 

Spectrograph

 field.

5.  Click the 

HoloSpec Setup

 button.

6.  Click 

Add

, then enter the “calibration coefficents” that can be 

found on the supplied Performance Sheet.

7.  Click the checkbox 

Enable calibration

, then click 

OK 

on the 

grating and spectrograph dialogs to exit the setup menu.

8.  Select your desired camera acquisition settings and take an acquisition – the calibration will be applied.

Configuring Solis for the HoloSpec

Camera Rear Panel Connections

Above: 

Rear panel shown for Newton camera. 

Refer to camera User Guide. Connections may 
vary with camera model.

Additional Software Functions (See the HoloSpec Hardware Guide for details of these procedures)

•  Small shifts in calibration can be accommodated using the pixel offset, found in the 

HoloSpec Setup

 dialog.

•  A new calibration can be calculated easily using a calibration lamp and the 

Calibrate/Manual X-Calibration

 dialog.

•  Units can be changed, e.g. from nanometers [nm] to Raman shift [cm

-1

], in the 

Calibrate/Change Units

 dialog.

The HoloSpec is supplied with camera, detector flange, 
grating and input slit pre-installed and aligned for ease of 
use

.

•  If you need to install a different slit or grating accessory please 

refer to the HoloSpec User Guide for instructions.

•  Do not adjust any lenses or components other than as 

described in the HoloSpec User Guide.

F/1.8i HoloSpec models

•  F/1.8i HoloSpec models are fitted with an additional 

holographic notch filter.

•  If you need to install a different notch filter please refer to the 

HoloSpec User Guide for instructions.

1. 

HoloSpec-F/1.8

 4. 

Detector flange

2. 

Grating 

 5. 

Input slit/adaptor

3. 

Detector (camera)

 6. 

Additional input  

 

 

 

    accessory

HoloSpec Components

Adjustment of Camera Focus

•  It is possible to adjust the focus of the camera, if required, 

using the camera focus thumbwheel on the HoloSpec.

Отзывы: