SPECIFICATIONS 

 

1.  Product name: Blood Pressure Monitor 

2.  Model: KD-5917 

3.  Classification:  Internally powered, Type  BF applied part,IPX0,No  AP  or APG,Continuous 

operation 

4.  Machine size: approx. 125mm×130mm×62mm (4 15/16

″ x 5 1/8″ x 2 7/16″) 

5.  Cuff circumference:  22cm-30cm(8 21/32

″ -11 13/16″), 30cm-42cm(11 13/16″ -16 17/32″) 

(Optional), 42cm-48cm(16 17/32

″ -18 29/32″) (Optional) 

6.  Weight: approx. 336g (11 27/32oz.) (exclude batteries) 

7.  Measuring method: Oscillometric method, automatic inflation and measurement 

8.  Memory volume: 60 times with time and date stamp 

9.  Power source: DC:6V

600mA, batteries: 4 ×1.5V

SIZE AA 

10.  Measurement range: 

Cuff pressure: 0-300mmHg 

Systolic:          60-260mmHg 

Diastolic:        40-199mmHg 

Pulse rate:      40-180 beats/minute 

11.  Accuracy: 

Pressure:        ±3mmHg 

Pulse rate:      ±5% 

12.  Environmental temperature for operation: 5

℃~

40

(41

℉~

104

13.  Environmental humidity for operation: 

90%RH 

14.  Environmental temperature for storage and transport: -20

℃~

55

(-4

℉~

131

15.  Environmental humidity for storage and transport: 

90%RH 

16.  Environmental pressure: 80kPa-105kPa 

17.  Battery life: Approx.180 times. 

18.  All  components  belonging  to  the  pressure  measuring  system,  including  accessories: 

Pump, Valve, LCD, Cuff and Sensor. 

 

Note: These specifications are subject to change without notice. 

 

NOTICE   
 

1.  Read all of the information in the operation guide and any other literature in the box before 

operating the unit. 

2.  Stay still, calm and rest for 5 minutes before blood pressure measurement. 

3.  The cuff should be placed at the same level as your heart. 

4.  During measurement, neither speak nor move your body and arm. 

5.  Measuring on same arm for each measurement. 

6.  Please always relax at least 1 or 1.5 minutes between measurements to allow the blood 

circulation in your arm to recover.

 

Prolonged over-inflation (

cuff pressure exceed 300 mmHg 

or maintained above15 mmHg for longer than 3 minutes

) of the bladder may cause ecchymoma 

of your arm. 

Содержание KD-5917

Страница 1: ...RIPTION 3 SPECIFICATIONS 4 NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CLOCK AND DATE ADJUSTMENT 6 3 CONNECTING THE CUFF TO THE MONITOR 7 4 APPLYING THE CUFF 7 5 BODY POSTURE DURING MEASUREMENT 7 6 TAKING YOUR BLOOD PRESSURE READING 8 7 DISPLAYING STORED RESULTS 8 8 DELETING MEASUREMENTS FROM THE MEMORY 9 9 ASSESSING HIGH BLOOD PRESSURE FOR ADULTS 9 10 TROUBLESHOOTING 1 10 11 T...

Страница 2: ...ht The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most people are awake and active Considering the above information it is recommended that you measure your blood pressure at approximately the same time each day Too frequent measurements may cause injury du...

Страница 3: ...d pulse rate can be measured automatically and non invasively The LCD display will show blood pressure and pulse rate The most recent 60 measurements can be stored in the memory with date and time stamp The Electronic Blood Pressure Monitor corresponds to the below standards IEC 60601 1 2005 EN 60601 1 2006 AC 2010 Medical electrical equipment Part 1 General requirements for basic safety and essen...

Страница 4: ...Environmental humidity for operation 90 RH 14 Environmental temperature for storage and transport 20 55 4 131 15 Environmental humidity for storage and transport 90 RH 16 Environmental pressure 80kPa 105kPa 17 Battery life Approx 180 times 18 All components belonging to the pressure measuring system including accessories Pump Valve LCD Cuff and Sensor Note These specifications are subject to chang...

Страница 5: ...that you consult with your physician for accurate assessment There are 2 conditions under which the signal of IHB will be displayed 1 The coefficient of variation CV of pulse period 25 2 The difference of adjacent pulse period 0 14s and the number of such pulse takes more than 53 percentage of the total number of pulse 13 Please do not use the cuff other than supplied by the manufacturer otherwise...

Страница 6: ...l replace all batteries with new ones Rechargeable batteries are not suitable for this monitor Remove the batteries if the monitor will not be used for a month or more to avoid relevant damage of battery leakage Avoid the battery fluid to get in your eyes If it should get in your eyes immediately rinse with plenty of clean water and contact a physician The monitor the batteries and the cuff must b...

Страница 7: ...it outward away from your body and tighten it and close the Velcro fastener b Place the cuff around a bare arm 1 2cm above the elbow joint c While seated place palm upside in front of you on a flat surface such as a desk or table Position the air tube in the middle of your arm in line with your middle finger d The cuff should fit comfortably yet snugly around your arm You should be able to insert ...

Страница 8: ...rily display the most recent one If no result has been stored zero will appear on LCD c Then the monitor inflates the cuff until sufficient pressure has built up for a measurement Then the monitor slowly releases air from the cuff and carries out the measurement Finally the blood pressure and pulse rate will be calculated and displayed on the LCD screen The blood pressure classification indicator ...

Страница 9: ...ter three beep Press the button MEM or START the monitor will turn off 9 ASSESSING HIGH BLOOD PRESSURE FOR ADULTS The following guidelines for assessing high blood pressure without regard to age or gender have been established by the World Health Organization WHO Please note that other factors e g diabetes obesity smoking etc need to be taken into consideration Consult with your physician for accu...

Страница 10: ...onitor 11 TROUBLESHOOTING 2 PROBLEM POSSIBLE CAUSE SOLUTION LCD Display shows EE The cuff was not properly applied or the rubber tube was bent or pressed Review the cuff applying and testing sections of the instructions and re test LCD Display shows abnormal result The cuff position was not correct or it was not properly tightened Apply the cuff correctly and try again Body posture was not correct...

Страница 11: ...riately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measurements or three years and the cuff integrity is maintained after 1 000 open close cycles of the closure 10 It is recommended the cuff should be disinfected 2 times every week if n...

Страница 12: ...ufacture s declaration electromagnetic emissions The KD 5917 is intended for use in the electromagnetic environment specified below The customer or the user of the KD 5917 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The KD 5917 uses RF energy only for its internal function Therefore its RF emissio...

Страница 13: ...e at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Guidance and manufacturer s declaration electromagnetic immunity The KD 5917 is intended for use in the electromag...

Страница 14: ...tion is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagneti...

Страница 15: ...eparation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the transmitter where...

Отзывы: