OPERATING MANUAL 

 

Continuous Level Sensor 

NSL-F 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SENSORS FOR FOOD AND LIFESCIENCE. 

NSL-F 

30002 / 2.0 / 2018-09-21 / YG / EU 

Содержание NSL -FR

Страница 1: ...OPERATING MANUAL Continuous Level Sensor NSL F SENSORS FOR FOOD AND LIFESCIENCE NSL F 30002 2 0 2018 09 21 YG EU...

Страница 2: ...nded use 3 2 Conformity with standards 3 3 Safety instructions 3 4 Special features advantages 4 5 Options accessories 4 6 Installation and connection 4 7 Installation of the Large User Interface LUI...

Страница 3: ...fied electricians or persons with technical training in electrics and by persons with the necessary authorization from the operator The power supply and the control circuit inputs and outputs must be...

Страница 4: ...12 coupling of nickel plated brass IP 67 shielded o M12 PVC 4G 5 m PVC cable 4 pin 5 m in length o M12 PVC 4G 10 m PVC cable 4 pin 10 m in length o M12 PVC 4G 25 m PVC cable 4 pin 25 m in length 6 Ins...

Страница 5: ...MPI software These contain additional useful information on changing the selected parameter 6 1 Setup using the MPI 200 programming adapter The MPI 200 programming adapter is connected to the NSL F c...

Страница 6: ...gnal no medium Sensor is not submerged in the medium Signal can be adjusted between 2 40 22 00 mA Parameter Node module Value name 4 mA Setpoint Signal Int x45a I Out 20 mA set value Signal Int x45a I...

Страница 7: ...o the next higher level Right or left button briefly R L scroll up or down Press and hold both buttons Press both buttons for 10 seconds then jump back to the beginning of the menu attention this is n...

Страница 8: ...all the way to the left with the left button until the field become gray Then press and hold the right button The system jumps to the selected parameter and it can be entered or changed in the same ma...

Страница 9: ...ut Signal Sel 001 to 254 330031 Underrange 2 40 mA 3 20 mA 3 40 mA 3 60 mA 3 80 mA 3 95 mA 4 00 mA 330141 Overrange 20 00 mA 20 05 mA 20 50 mA 21 00 mA 21 20 mA 21 40 mA 21 60 mA 21 80 mA 22 00 mA 330...

Страница 10: ...nction State Src Discrimin Threshold 331112 Direction Norm Open Norm Clsd 331122 Inp Method SP Hyst 2Pt MinMax 331132 Switch Pnt 025 0 bis 125 0 in 0 1er Steps 331142 Hysteresis 00 0 bis 50 0 in 0 1er...

Страница 11: ...erdrive limit 3 4 mA set value 4 20 mA set value 5 Overdrive limit 6 Error signal Value overflow Warning signal no medium Sensor is not submerged in the Medium Signal can be set between 2 40 22 00 mA...

Страница 12: ...installed to activate the display 7 1 Retrofitting the display if a display was not yet installed After removing the plastic cover the large display can be inserted in the head of the sensor The tabs...

Страница 13: ...ck with the display out of the sensor housing and pull the small display off of the puck Now the puck without the display can be properly positioned and installed back in the sensor head and the wirin...

Страница 14: ...NSL F Series Operating Manual 14...

Страница 15: ...NSL F Series Operating Manual 15 8 Dimensions...

Страница 16: ...ing figure only contains information on the rod design The sizes of the head and process con nection can be gathered from the drawings for NSL F 00 NSL F 02 Except for the rod length of 200 mm up to m...

Страница 17: ...NSL F Series Operating Manual 17 NSL FR Remote Version...

Страница 18: ...tage Electrical connection with 5 PIN M12 plug connection option L Anderson Negele 5 PIN Cordset L Standard 5 PIN Cordsets Description 1 Red 1 Brown Supply voltage 2 Black 2 White Supply voltage 3 Gre...

Страница 19: ...177 2415 Stainless steel 1 4404 Ra 0 8 m Temperature ranges Environment Storage temperature Process CIP SIP 0 70 C 40 85 C 10 140 C 143 C max 120 min Resolution Rod length 500 mm Rod length 500 mm 0...

Страница 20: ...nik GmbH Raiffeisenweg 7 87743 Egg an der Guenz Germany Phone 49 83 33 92 04 0 Fax 49 83 33 92 04 49 Anderson Instrument Company Inc 156 Auriesville Road Fultonville NY 12072 Phone 1 518 922 5315 Fax...