Installation Guidelines
Quick
release
connector
2" PVC pipe
Ceiling
Adapter (1)
Quick
release
connector
Quick
release
connector
Exterior*
Roof
Exterior*
Ex
te
rior*
Recommended Installation
Alternate Option 1
Roof installation with vent boot
2" PVC pipes
When possible, the Anprolene Sterilizer
should be located near an outside wall,
and exhaust directly to the outside by
means of ¾" (1.9cm) I.D. tubing.
Through wall mounting kits are available
from Andersen. Note: Sterilizer must be
placed in an area that can maintain 68°F
or above room temperature.
Through the roof applications are best accomplished by installing 2" SCH
40 PVC using a roof vent boot and stabilizing the PVC to extend just below
the ceiling. The ¾" I.D. to 2" adapter cap is available from Andersen.
Note: 2” SCH 40 PVC exhaust can run up to 800' depending on path -
bends/90° turns will reduce distance.
Anprolene sterilizer installation options
Hose