![AND Weighing Indicator AD-4407 Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/and/weighing-indicator-ad-4407/weighing-indicator-ad-4407_instruction-manual_2941327050.webp)
Page
48
AD-4407 Weighing Indicator
When the memory is short to store the data of UFC commands, an “M” is sent back.
Optional addresses can be appended to a command. The address form is
"@address" and the address is specified at F-Function
f06
. The reply (data or
error code) is also sent with the address.
Example: Command is "Display net value". Address is 23.
Sending a command
Sending a command with address
Unperformed command reply
Unknown command reply
Memory overflow (for UFC commands)
13.3.1.
Command to Request Data
Request Displ
Request Displ
Request Displ
Request Display Data (1)
ay Data (1)
ay Data (1)
ay Data (1)
When receiving this command, returns the displayed data immediately.
Template
RW
Command
Reply
Request Display Data (2)
Request Display Data (2)
Request Display Data (2)
Request Display Data (2)
When receiving this command, returns the displayed data immediately with format 1
or 2.
Template
RW,1 or RW,2
Command
or
Reply
Format1(
f34
) or
Format2(
f35
)
Request Gross weight
Request Gross weight
Request Gross weight
Request Gross weight
When receiving this command, returns the gross data immediately.
Template
RG
Command
Reply
13.3. Command Format
W
R
G
R
W
R
1
,
W
R
2
,
T
S
G
,
,
S
0
0
+
2
1
.
3
k
0
g
T
S
G
,
,
S
0
0
+
2
1
.
3
k
0
g
M
Terminators
Memory overflow error
Carriage return(0Dh)
Line feed(0Ah)
I
Unperformed command
Terminators
N
M
Terminators
Command
Terminators
Command
N
M
@ 2 3
Address number
Unknown command
Terminators
?
@ 2 3
Address number