The C.P.G. Indicator 

 

The C.P.G. Indicator

 

 

The C.P.G. (

C

orrect 

P

osition 

G

uidance) indicator is the function to inform a 

difference between the height (wrist angle) of the blood pressure monitor and 
your cardiac height in the correct posture (Example: sitting posture, height of 
table and chair, etc.) during the measurement. The indicator can be used to 
get more stable measurement condition. 

 

The C.P.G. Indicator 

 

 

The height of the blood pressure 
monitor is lower than your heart. 

 
Low angle 

Incorrect Height

 

 
 
 
 

The height of the blood pressure 
monitor is the same level as your 
heart. 

Proper angle 

Correct Height

 

 
 

The height of the blood pressure 
monitor is higher than your heart. 

 
High angle 

Incorrect Height

 

 

 

The position of the device is checked both before and after measurement If both 
checks show a correct measurement position the LEVEL indicator is lit(blue). 
For all other measurements an indicator for LOW or HIGH measurement 
position will be lit (orange). 

 

How To Use My C.P.G.

   

 

The C.P.G. function can be used with proper posture (wrist angle) in the majority 
of measurements. If you need to change the posture to adjust the height so that 
the height of the blood pressure monitor is the same level as your cardiac 
height, you can use my C.P.G. function to store a personal posture. Preset your 
angle to my C.P.G. function before measurement. 

Proper Posture

 

Sit straight back on a 

chair and put your 

elbow on a table. 

Table

Chair

 

  The

C.P.G. indicator 

The

 C.P.G. indicator : 

LOW

 

 

The

C.P.G. indicator : 

LEVEL

 

The

C.P.G. indicator : 

HIGH

Содержание UB-533

Страница 1: ...Wrist Digital Blood Pressure Monitor Model UB 533 Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione 使用手冊 1WMPD4003121E 翻譯 ...

Страница 2: ...P G and my C P G 11 8 Measurement 12 9 After Measurement 12 Measurements 13 Recalling the Memory Data 14 Deleting Data Stored in Memory 15 What Is The IHB AFib Indicator 15 What Is The AFib 15 IHB AFib 16 Pressure Bar Indicator 17 WHO Classification Indicator 17 The C P G Indicator 18 About Blood Pressure 19 What is Blood Pressure 19 What is Hypertension and How is it Controlled 19 Why Measure Blo...

Страница 3: ...ould be avoided Clean the device with a dry soft cloth or a cloth dampened with water and a neutral detergent Never use alcohol benzene thinner or other harsh chemicals to clean the device Avoid tightly folding the cuff for long periods as such treatment may shorten the life of the components The device is not water resistant Prevent rain sweat and water from soiling the device Measurements may be...

Страница 4: ...ear cuff may become hot and potentially cause malfunction Use of accessories not detailed in this manual may compromise safety Should the battery short circuit it may become hot and potentially cause burns Allow the device to adapt to the surrounding environment before use about one hour Do not inflate without wrapping the cuff around your wrist Contraindications The following are precautions for ...

Страница 5: ...Date clock display AM PM mark Alkaline batteries LR03 or AAA Battery cover button average memory and change of preset parameters SET button user and clock START button Display Heart mark User and User Battery indicator MEMORY Average WHO classification indicator and Pressure bar indicator Cuff Cuff Fit Error Symbol Movement Error Symbol IHB AFib ...

Страница 6: ...ery installation guide Direct current SN Serial number Date of manufacture Type BF Device and cuff are designed to provide special protection against electrical shocks EC directive medical device label IP International protection symbol WEEE label Manufacturer EU representative Refer to instruction manual booklet Keep dry ...

Страница 7: ...very slight vibration like shivering or shaking is detected Appears when a body or arm movement is detected The reading may yield an incorrect value Take another measurement Remain still during measurement Appears during measurement when the cuff is attached loosely The reading may yield an incorrect value Apply the cuff correctly and take another measurement Deleted rate of IHB AFib in memory Num...

Страница 8: ...l during measurement The systolic and diastolic values are within 10 mmHg of each other Apply the cuff correctly and take another measurement The pressure value did not increase during inflation The cuff is not applied correctly PUL DISPLAY ERROR The pulse is not detected correctly Blood pressure monitor internal error Remove the batteries and press the START button and then install the batteries ...

Страница 9: ...rten the battery life or cause the device to malfunction LOW BATTERY mark does not appear when the batteries are drained The battery life varies with the ambient temperature and may be shorter at low temperatures Generally two new LR03 AAA batteries will last approximately for three months when used twice for measurement each day Use the specified batteries only The batteries provided with the dev...

Страница 10: ...ing the button Press the SET button to set the current day and move to hour minute selection 5 Select the hour using the button Press the SET button to set the current hour and move to minute selection 6 Select the minute using the button Press the SET button to proceed to 6 The C P G Function Switch Holding down the button will change the value continuously Pressing the START button will turn the...

Страница 11: ...about the same time every day An individual s blood pressure varies constantly depending on what they are doing what they have eaten and what they drink can have a very strong and rapid effect on your blood pressure Do not measure immediately after physical exercise or a bath Rest for twenty or thirty minutes before taking the measurement Do not cross your legs Keep your feet flat on the floor and...

Страница 12: ...onitor to the same level as your heart using your wrist angle 2 Select an indicator using the button Indicator The C P G function is used my C P G function OFF Data of my C P G is deleted Proceed to step 3 Indicator my C P G function is ON and the current angle is stored when switched to the indicator Proceed to step 3 3 Press the START button to turn off the device Note When removing the batterie...

Страница 13: ...stored in memory While readings are displayed if you press the button to turn the device off new readings are not stored Remove the cuff and record your data Notes The device is provided with an automatic power shut off function which stores the current data in the memory and turns the device off automatically one minute after measurement Allow at least three minutes between measurements on the sa...

Страница 14: ...and the heart mark blinks Note If you wish to stop inflation at any time press the START button again 5 When the measurement is complete the device displays readings of the systolic and diastolic pressure readings pulse rate WHO classification I H B symbol and C P G indicator While readings are displayed the date and time are displayed alternately The cuff exhausts the remaining air and completely...

Страница 15: ... all PM evening measurements taken between 18 00 and 1 59 Data of number and measurement data The device displays in order from most recent data The date and time are displayed alternately while displaying the measurement data In the example No 30 data No 29 data No 01 data 3 If you press the button after oldest data is displayed the device proceeds to step 1 the average of all measurements and th...

Страница 16: ...ular rhythm during the measurements the IHB AFib indicator will appear on the display with the measurement values Note We recommend contacting your physician if you see this IHB AFib indicator frequently What Is The AFib The heart contracts due to electrical signals occurring in heart and sends blood through the body Atrial fibrillation AFib occurs when the electrical signal in the atrium becomes ...

Страница 17: ...mend contacting your physician if IHB AFib level is high Display of IHB AFib IHB AFib is displayed when displaying average values IHB AFib is not displayed when the memory number is six or less Level 0 IHB AFib 0 Level 1 IHB AFib 1 9 Level 2 IHB AFib 10 24 Level 3 IHB AFib 25 100 Not displayed IHB AFib Average value display Number of detected IHB AFib s in memory x 100 Total number ...

Страница 18: ...ressure classification is described on 20 page Example Moderate hypertension Mild hypertension High normal The indicator displays a segment based on the current data corresponding to the WHO classification WHO Classification Indicator Severe hypertension Moderate hypertension Mild hypertension High normal Normal Optimal Pressurizing Inflation in progress Measurement in progress Inflation complete ...

Страница 19: ...r than your heart High angle Incorrect Height The position of the device is checked both before and after measurement If both checks show a correct measurement position the LEVEL indicator is lit blue For all other measurements an indicator for LOW or HIGH measurement position will be lit orange How To Use My C P G The C P G function can be used with proper posture wrist angle in the majority of m...

Страница 20: ...fore eating What is Hypertension and How is it Controlled Hypertension an abnormally high arterial blood pressure if left unattended can cause many health problems including stroke and heart attack Hypertension can be controlled by altering lifestyle avoiding stress and with medication under a doctor s supervision To prevent hypertension or keep it under control Do not smoke Exercise regularly Red...

Страница 21: ...ng the day In hypertensive individuals variations are even more pronounced Normally the blood pressure rises while at work or play and falls to its lowest levels during sleep So do not be overly concerned by the results of one measurement Take measurements at the same time every day using the procedure described in this manual to get to know your normal blood pressure Regular readings give a more ...

Страница 22: ...ed properly Apply the cuff correctly You moved your wrist or body during the measurement Make sure you remain very still and quiet during the measurement The cuff position is not correct Sit comfortably and still Place your arm on a table with your palm facing upward and the cuff at the same level as your heart If you have a very weak or irregular heat beat the device may have difficulty in determ...

Страница 23: ...spected every 2 years to ensure proper functioning and accuracy Please contact the authorized dealer in your area or A D for maintenance Technical Data Type UB 533 Measurement method Oscillometric measurement Measurement range Pressure 0 299 mmHg Systolic pressure 60 279 mmHg Diastolic pressure 40 200 mmHg Pulse 40 180 beats minute Measurement accuracy Pressure 3 mmHg Pulse 5 Power supply 2 x 1 5 ...

Страница 24: ...ns 20 to 60 C 10 to 95 RH 700 to 1060 hPa Dimensions Approx 56 W x 88 H x 21 5 D mm Weight Approx 95 g excluding the batteries Ingress protection IP20 Note Specifications are subject to change for improvement without prior notice IP classification is the degrees of protection provided by enclosures in accordance with IEC 60529 This device is protected against solid foreign objects of 12 mm diamete...

Страница 25: ...English 24 MEMO ...

Страница 26: ... 263 0119 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 OOO A D RUS OOO ЭЙ энд ДИ РУС 121357 Российская Федерация г Москва ул Верейская дом 17 Business Center Vereyskaya Plaza 2 121357 Russian Federation Moscow Vereyskaya Street 17 тел 7 495 937 33 44 факс 7 495 937 55 66 A D Technology Trading Shanghai Co Ltd 爱安德技研贸易 上海 ...

Отзывы: