When using the extension cable, the temperature characteristics and noise immunity of the
scale may deteriorate.
You must calibrate the scale after the extension cable has been connected.
(Refer to
“12. Calibration” in the HV-CWP/HW-CWP Series Instruction Manual for details.)
AX-KO6430-5M
Extension Cable (
6, shield cable 5m)
AX-TMBOXTS-6
Junction Box (IP67)
(AX-043005266)
Indicator Holder
1. Remove the load cell cable from the indicator following the steps in “5.2. Removing the
Pole” in the HV-CWP/HW-CWP Series Instruction Manual.
2. Open the cover of the junction box.
3. Pass the cable through the hole of the cap, waterproof rubber sleeve and junction box.
Use the following waterproof rubber sleeves:
For load cell cable: Sleeve A
For extension cable: Sleeve B
Sleeve A Sleeve B
Cover
Cap
Waterproof Rubber Sleeve
Terminal Block
Buttons
HV-CWP/HW-CWP Series How to Extend Load Cell Cable
Connecting the Extension Cable
Precautions
Extension Cable
Indicator Base Unit
Junction Box
(Indicator Holder)
Required Items