When  using  the  extension  cable,  the  temperature  characteristics  and  noise  immunity  of  the 
scale may deteriorate. 
You must calibrate the scale after the extension cable has been connected. 
(Refer to 

“12. Calibration” in the HV-CWP/HW-CWP Series Instruction Manual for details.) 

 

 

 

  AX-KO6430-5M 

    Extension Cable (

6, shield cable 5m) 

  AX-TMBOXTS-6 

    Junction Box (IP67) 

  (AX-043005266) 

    Indicator Holder 

 

 

 

 

 

 

 

 

 

 

1.  Remove  the  load  cell  cable  from  the  indicator  following  the  steps  in  “5.2.  Removing  the 

Pole” in the HV-CWP/HW-CWP Series Instruction Manual. 

2.  Open the cover of the junction box. 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.  Pass the cable through the hole of the cap, waterproof rubber sleeve and junction box. 

Use the following waterproof rubber sleeves:

 

  For load cell cable: Sleeve A 

  For extension cable: Sleeve B 

 

 

Sleeve A                  Sleeve B 

Cover 

Cap 

Waterproof Rubber Sleeve 

Terminal Block 

Buttons 

HV-CWP/HW-CWP Series    How to Extend Load Cell Cable 

Connecting the Extension Cable 

Precautions 

Extension Cable 

Indicator                                                                Base Unit 

Junction Box 

(Indicator Holder) 

Required Items 