AND BA-125DTE Скачать руководство пользователя страница 16

16 

Display 

Error code 

Description and possible countermeasure 

Weighing 
display 
-E 

 

Weighing pan error 

The mass value is too light. Check that the weighing pan is installed 
correctly. Set the weighing pan correctly. Calibrate the balance. 

rtc PF 

 

Clock battery error 

The clock backup battery has been depleted. Set the time and date. 
Even if the clock backup battery is depleted, the clock and calendar 
function works normally as long as the balance is powered with the 
AC adapter. If this error appears frequently, contact your local dealer. 

LoWVoLt 

 

Power supply voltage fault 

The voltage supplied from the AC adapter is abnormal. 
Please check if the problem is the AC adapter supplied with the 
balance. 

SD Error 
 
MW Error 

 

Repeatability error 

With the minimum weighing warning function, the standard deviation 
(SD) of repeatability exceeded 50 digits. Review the installation 
environment of the balance. For details, refer to the separate 
instruction manual listed in “1.1. Detailed Manuals”. 
“SD Error” appears in repeatability display. 
“MW Error” appears in minimum weighing value (reference value). 

 

EC

E00

 

Communications error 

A protocol error occurred in communications. Check the format, baud 
rate, and parity. 

 

EC

E01

 

Undefined command error 

An undefined command was found. Check the transmitted command. 

 

EC

E02

 

Not ready 

The received command cannot be executed: 
(e.g.) Q command was received when not in weighing mode. 
(e.g.) Q command was received while rezeroing. 
Adjust the delay time to transmit a command. 

 

EC

E03

 

Timeout error 

With the command timeout setting, there was a wait time of 
approximately 1 second or more while receiving command 
characters. Check the communication. 

 

EC

E04

 

Character length error 

The number of characters in the received command has exceeded 
the limit. Check the command to transmit. 

 

EC

E06

 

Format error 

The description of the received command is incorrect: 
(e.g.) The number of digits of numerical values is incorrect. 
(e.g.) There are alphabet characters among the numerical values. 
Check the transmitted command. 

 

EC

E07

 

Parameter setting error 

The value of the received command has exceeded the allowed value. 
Check the setting range of the numerical value of the command. 

 

Содержание BA-125DTE

Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004296 ...

Страница 2: ...lated into any language in any form by any means without the written permission of A D Company Ltd The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice Windows Word and Excel are registered trademarks of the Microsoft Corporation ...

Страница 3: ...en Buttons and Key Panel Basic Operation 10 5 IR sensors and Automatic Doors 12 5 1 IR sensors 12 5 2 Automatic doors 12 6 Check Function 13 7 Sensitivity Adjustment Calibration Test 14 7 1 Automatic sensitivity adjustment 14 7 2 Sensitivity adjustment using the internal mass 14 8 Typical Errors Error Codes 15 9 Ionizer AX ION 25 17 9 1 Using the Ionizer 17 10 Specifications 18 11 Compliance 19 11...

Страница 4: ...ly and Installation Caution A BA TE series analytical balance is a precision instrument and it should be carefully unpacked It is advisable to store the packing materials so that they can be used when transporting the balance The contents of the package vary depending on the product Refer to the illustration of the packing contents and make sure that everything is included Do not connect the AC ad...

Страница 5: ...pane Base stand Aluminum round pans for analysis 10 pieces each Large pan Φ15 0 8 mL Medium pan Φ12 0 3 mL Breeze break doors Small pan Φ8 0 05 mL Pan A Φ25 Breeze break ring A Dust plate B Breeze break front glass Breeze break doors Breeze break unit Weighing unit Display unit protection cover Breeze break top door Breeze break bottom plate Display unit Tweezers Cable A Cable B Cable C USB cable ...

Страница 6: ...emble the five parts of the weighing pan in the correct position by referring to the cross sectional view For BA 225TE BA 225DTE BA 125DTE assemble the four parts of the weighing pan in the correct position by referring to the cross sectional view Breeze break unit Step 1 Step 3 Step 2 Weighing unit For BA 225TE BA 225DTE BA 125DTE For BA 6TE BA 6DTE Pan A Pan support A Breeze break ring A Dust pl...

Страница 7: ...the left and right sides 7a Insert the breeze break doors on the left and right sides 7b Secure the breeze break doors with the latches Step 8 Assemble the breeze break front glass 8a Insert the breeze break front glass 8b Secure the breeze break front glass with the latches Notch Example of the small breeze break with the right side open Base stand Step 5 Top pane Side panes 3 pieces Base stand S...

Страница 8: ...irection of the arrow on each cable Step 10 Connect the breeze break unit and ionizer with the included cable C Caution The ionizer operates when the AC adapter is connected to the balance Step 11 Connect the included AC adapter to the balance BA 225TE BA 225DTE BA 125DTE BA 6TE BA 6DTE Breeze break unit Display unit Weighing unit Ionizer Step 9 Step 11 AC adapter Step 10 ...

Страница 9: ... for at least an hour or at least four hours for BA 6TE BA 6DTE with the AC adapter connected to the power supply Perform sensitivity adjustment of the balance before using it for the first time or after having moved it to another location so that accurate weighing can be performed For details refer to the detailed instruction manual listed in 1 1 Detailed Manuals Caution Do not install the balanc...

Страница 10: ... P TEST button 2 Application button 10 Communication device setting button 3 Left IR sensor function change button 11 Right IR sensor function change button 12 Right IR sensor 4 Left IR sensor 5 Stabilization indicator 6 ON OFF key 14 LOG IN key 7 HOME key 15 MENU key 8 PRINT button 16 RE ZERO button 13 Alert ...

Страница 11: ...refer to the separate instruction manual listed in 1 1 Detailed Manuals 9 CAL button Displays the sensitivity adjustment calibration screen For details refer to the separate instruction manual listed in 1 1 Detailed Manuals 10 Communication device setting button Displays the communication device setting screen Configures the settings of the connected communication device 11 Right IR sensor functio...

Страница 12: ...the breeze break doors are assigned to and for the doors to be fully open or partially open in the function table of the balance For details refer to the separate instruction manual listed in 1 1 Detailed Manuals Opening the breeze break door Step 1 If you want to open the breeze break door on the left side or on the right side when it is closed hold your hand over the right IR sensor or the left ...

Страница 13: ... Additionionally whether or not the balance s performance is being exhibited can be easily checked by checking the repeatability with the internal mass The check function can be selected from the menu screen For details refer to the separate instruction manual listed in 1 1 Detailed Manuals ...

Страница 14: ...tions are displayed on the screen for how to adjust the sensitivity 2 Internal sensitivity adjustment The internal mass is used to automatically adjust the balance 3 Internal calibration test The internal mass is used to check the accuracy of weighing and the result is output Note that the balance is not adjusted 4 External calibration test Your external calibration weight is used to check the acc...

Страница 15: ... Error 8 Abnormality in the internal memory data of the balance If this error continues to be displayed repair is necessary Please contact your local dealer for repair Error 9 Abnormality in the internal memory data of the balance If this error continues to be displayed repair is necessary Please contact your local dealer for repair Con Err Weighing unit error Data from the weighing unit has not b...

Страница 16: ...or appears in repeatability display MW Error appears in minimum weighing value reference value EC E00 Communications error A protocol error occurred in communications Check the format baud rate and parity EC E01 Undefined command error An undefined command was found Check the transmitted command EC E02 Not ready The received command cannot be executed e g Q command was received when not in weighin...

Страница 17: ... it in and then turn on the power The power lamp of the ionizer lights up Step 3 Place the target object within the effective range of static elimination Step 4 Hold your hand over the IR sensor touchless infrared proximity sensor on the front of the ionizer to start static elimination The ION lamp indicating that static elimination is in progress lights up Step 5 Refer to the figure below for the...

Страница 18: ...TE BA 125DTE Weighing capacity 220 g 220 g 120 g 51 g 51 g Maximum display 220 00084 g 220 0008 g 120 0008 g 51 00009 g 51 00009 g Readability 0 01 mg 0 1 mg 0 1 mg 0 01 mg 0 01 mg Repeatability Standard deviation 0 015 mg 0 03 mg 50 g 200 g 0 1 mg 200 g 0 1 mg 100 g 0 025 mg 50 g 0 030 mg 50 g Linearity 0 15 mg 0 20 mg 0 20 mg Display 5 inch WVGA TFT LCD color display Operation Touchscreen and ke...

Страница 19: ...s These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is li...

Страница 20: ...MEMO ...

Страница 21: ...MEMO ...

Страница 22: ...MEMO ...

Страница 23: ......

Страница 24: ...treet Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 한국에이 엔 디 주 서울특별시 영등포구 국제금융로6길33 여의도동 맨하탄빌딩 817 우편 번호 07331 817 Manhattan Bldg 33 Gukjegeumyung ro 6 gil Yeongdeungpo gu Seoul 07331 Korea 전화 82 2 780 4101 팩스 82 2 782 4264 OOO A D RUS OOO ЭЙ энд ДИ РУС 121357 Российская Федерация г Москва ул Верейская дом 17 Business Center Vereyskaya Plaza...

Отзывы: