16
Display
Error code
Description and possible countermeasure
Weighing
display
-E
Weighing pan error
The mass value is too light. Check that the weighing pan is installed
correctly. Set the weighing pan correctly. Calibrate the balance.
rtc PF
Clock battery error
The clock backup battery has been depleted. Set the time and date.
Even if the clock backup battery is depleted, the clock and calendar
function works normally as long as the balance is powered with the
AC adapter. If this error appears frequently, contact your local dealer.
LoWVoLt
Power supply voltage fault
The voltage supplied from the AC adapter is abnormal.
Please check if the problem is the AC adapter supplied with the
balance.
SD Error
MW Error
Repeatability error
With the minimum weighing warning function, the standard deviation
(SD) of repeatability exceeded 50 digits. Review the installation
environment of the balance. For details, refer to the separate
instruction manual listed in “1.1. Detailed Manuals”.
“SD Error” appears in repeatability display.
“MW Error” appears in minimum weighing value (reference value).
EC
,
E00
Communications error
A protocol error occurred in communications. Check the format, baud
rate, and parity.
EC
,
E01
Undefined command error
An undefined command was found. Check the transmitted command.
EC
,
E02
Not ready
The received command cannot be executed:
(e.g.) Q command was received when not in weighing mode.
(e.g.) Q command was received while rezeroing.
Adjust the delay time to transmit a command.
EC
,
E03
Timeout error
With the command timeout setting, there was a wait time of
approximately 1 second or more while receiving command
characters. Check the communication.
EC
,
E04
Character length error
The number of characters in the received command has exceeded
the limit. Check the command to transmit.
EC
,
E06
Format error
The description of the received command is incorrect:
(e.g.) The number of digits of numerical values is incorrect.
(e.g.) There are alphabet characters among the numerical values.
Check the transmitted command.
EC
,
E07
Parameter setting error
The value of the received command has exceeded the allowed value.
Check the setting range of the numerical value of the command.