Analytik Jena UVP iBox Explorer2 Скачать руководство пользователя страница 6

iBox Explorer

2

  

 

Specifications 

Power Requirements: 

100-230V; 6.3 Amps at 120 Volts       

Mains supply voltage fluctuations are not to exceed 10 percent of the nominal supply 
voltage 

 

 

Pollution Degree:  

Installation Category:  

II 

Altitude:  

Up to 2000m 

Ambient Temperature:  

5°C to 40°C 

Software Requirements: 

Windows XP (SP2) or Windows 7 (32-bit and 64-bit versions) 

Humidity:  

Maximum relative humidity of 80% for temperatures up to 31°C, decreasing linearly  
to 50% maximum relative humidity at 40°C 

 

Darkroom Cabinet  

The iBox Explorer

2

 is constructed of aluminum and fabricated to provide a light tight chamber. Darkroom 

dimensions are 17.5”W x 20”D x 42”H (44.5cm x 50.8cm x 106.7cm). 

Four Position Filter Tray 

The filter tray accommodates up to four removable emission filters. The iBox Explorer

system includes RFP 

(Red Fluorescent Protein), GFP (Green Fluorescent Protein), 515 Long Pass and Neutral Density emission 
filters.  VisionWorksLS software controls the selection of filters. Additional filters are available through UVP. 

 

CCD Camera and Objectives 

The high-sensitivity CCD camera and fast optics, housed in the top of darkroom, generate high resolution 
images. The camera and motorized optics are controlled by VisionWorksLS software. 

Magnification and Field of View 

The iBox Explorer

2

 contains a combination of relay and objective lenses, providing magnifications of 0.17x, 

0.25x, 0.50x, 1.66x, 2.50x, 4.50x, 7.50x, 8.80x and 16.5x. The corresponding fields of view are 90mm

2

, 60mm

2

30mm

2

, 9mm

2

, 6mm

2

, 3.3mm

2

, 2mm

2

, 1.7mm

2

 and 0.9mm

2

Viewer Window 

 

The viewer window, built into the darkroom door, opens for sample viewing.  

 

Motorized Sample Platform 

The motorized sample platform, controlled by the joystick, can be adjusted in the X, Y and Z dimensions.  The 
platform can be moved 100mm in each direction.   

Warming Plate  

 

The warming plate creates a constant temperature of 38ºC for the warming of animal specimens. The dark 
background creates an ideal imaging surface while maintaining a constant temperature for the animal. 

 

VisionWorksLS Software 

The VisionWorksLS Acquisition & Analysis software controls the darkroom functions and lighting as well as the 
motorized optics and camera. The software is operational on both 32-bit and 64-bit Windows operating systems. 

 

BioLite

 

Xe MultiSpectral Source 

The BioLite Xe MultiSpectral Source provides dual path excitation lighting for specimens. GFP and RFP filters 
come standard with the BioLite Xe, and custom filters for specific applications are also available. Refer to 
“Replacement Parts and Accessories” for ordering information. 

Содержание UVP iBox Explorer2

Страница 1: ...________________________________ UVP LLC 2066 W 11th Street Upland CA 91786 Tel 909 946 3197 800 452 6788 Fax 909 946 3597 Web Site www uvp com 81 0342 01 Rev E Ultra Violet Products Ltd Unit 1 Trinit...

Страница 2: ...e Xe MultiSpectral Source 6 Installation 7 Installing the Software 7 Registering the Software 7 Installing the Hardware 9 BioLite Xe MultiSpectral Light Source Setup 10 Fiber Optic Cable Connection 11...

Страница 3: ...ed imaging platform The system comes with dual path directed lighting for exciting samples with the BioLite Xe MultiSpectral Light Source The BioLite Xe offers an eight position filter wheel six posit...

Страница 4: ...in a laboratory environment by personnel knowledgeable in safe laboratory practices Proper precautions must be taken as potential electrical and mechanical hazards are inherent in the use of this equi...

Страница 5: ...lip for a complete equipment list System components may vary Note The system may also include a computer and monitor Monochrome CCD camera and magnification objectives are housed inside the darkroom F...

Страница 6: ...tics are controlled by VisionWorksLS software Magnification and Field of View The iBox Explorer 2 contains a combination of relay and objective lenses providing magnifications of 0 17x 0 25x 0 50x 1 6...

Страница 7: ...on is required To immediately activate the software online choose On the Fly activation If the computer is not connected to the Internet select Offline activation and proceed to the following page of...

Страница 8: ...to the Internet click Offline activation to register the software This allows the user to obtain the activation code and enter it at another time Click Next to continue Click the link provided and com...

Страница 9: ...ositions as noted below Once all filters are in place position the magnetic filter holder over the filter tray to hold the filters in place Open the darkroom door and slide the filter tray well door o...

Страница 10: ...d The xenon bulb assembly will come separate from the BioLite Xe and is equipped with a male connector on one end and a female connector on the other CAUTION Wear safety glasses when handling the xeno...

Страница 11: ...e the filter assembly inside the BioLite Xe wheel slot If the filter assembly does not easily insert check to ensure that the orientation of the assembly is correct Fiber Optic Cable Connection The iB...

Страница 12: ...E Allow the warming plate to reach room temperature prior to operating If the unit is extremely cold when switched on the warming surface temperature can exceed the preset temperature indicated on the...

Страница 13: ...VisionWorksLS software Turn the darkroom ON if not already powered on power switch is located on the lower back side of the darkroom Install the camera driver o When the camera is first connected to t...

Страница 14: ...te the darkroom controls click on the Acquisition tab in VisionWorksLS The software operates the objectives lenses emission filters and BioLite Xe lighting The purpose of the Action Tabs is to enable...

Страница 15: ...Lighting VisionWorksLS provides automated control of the BioLite Xe excitation filters and lighting via the Lighting menu Note that it is possible to control the BioLite Xe manually via the controls...

Страница 16: ...at a higher magnification then move to a lower magnification without losing focus The iBox Explorer 2 s various magnification settings provide the following fields of view Optical Magnification 0 17x...

Страница 17: ...w button under the Bookmarks section of the Microscope tab Enter a name for the bookmark for reference New allows the user to save a new bookmark Update allows the user to update the name or location...

Страница 18: ...t the desired intensity Adjust the position of the stage using the joystick as described in Adjusting the Sample Platform Adjust the exposure time binning and region of interest ROI while in the Previ...

Страница 19: ...ng Return Procedure A Returned Goods Authorization RGA number must be obtained from UVP Customer Service before returning any product or component Troubleshooting No Power to the Darkroom 1 Recheck th...

Страница 20: ...cale Mitsubishi 89 0069 06 115V Thermal printer digital archive quality 256 grayscale Mitsubishi 89 0069 07 230V Thermal printer digital archive quality 256 grayscale Sony for Europe 89 0069 15 230V T...

Страница 21: ...52 6788 or 909 946 3197 and ask for Technical Support during regular business days between 7 00 am and 5 00 pm PST Call 44 0 1223 420022 and ask for Customer Service during regular business days betwe...