Specification
40
Sensitivity
1 nmol/l FAM at 30 µl
Sample volume
1 nmol/l FAM at 10 µl
Sample volume
Measuring time
approx. 6 s for a simple measurement, 6 colors
Measurement range
± 130 000 (±17 bit)
Dynamic range
10 log stages
Measuring principle
Fiber-optic shuttle system
with 8 optical fibers
Fiber-optic shuttle system
with 16 optical fibers
Light source
Four-color high-performance LEDs (blue, green, white, red)
Detector
Highly-sensitive PMT (Photo Multiplier Tube)
Optimum signal/noise characteristics due to efficient noise
suppression
Filter modules
12 color-, FRET- and protein modules
6 positions in the device
Safety features
Sample chamber is monitored by sensors
Overtemperature protection in the heated lid
Interlock switch (front flap)
Noise emissions
< 45 dB
Regulations/certificates
See “Standards and directives” on page 42
10.1.2
Power module
Designation/type
qTOWER
3
auto power module
Dimensions
(width x height x depth)
290 x 186 x 335 mm
Mass
approx. 12 kg
Protection type
IP 20
Safety class
I
Voltage range*
100 / 115 / 230 V
Max. power
950 W
Frequency
50/60 Hz
Fuse
2x 10 AT / 250 V
Output ports
3 (Power, Data 1, Data 2)
Safety features
Overload detection
Ground fault
Overheating detection
Sheathed connectors and sockets
Regulations/certificates
See “Standards and directives” on page 42
qPCR
application
Fluorescence
spectrometer
Other
characteristics
General
data
Electrical
characteristics
Other
characteristics