Specification
16
Edition 05/2014
PlasmaQuant PQ 9000
3
Specification
3.1
Technical data
3.1.1
PlasmaQuant PQ 9000 data
Optical system
Monochromator
Echelle grating double-monochromator with a focal length of F=380 mm and variable
gap; pre-monochromator with quartz prism, wavelength selection via additional
reflected neon radiator
Wavelength range
160 – 900 nm
Wavelength accuracy
< 0.4 pm
Spectral resolution
0.002 nm at 2 nm
Grating
Mechanically ruled grating, 79 grooves/mm, blaze angle of 74°r
Optical bank
encapsulation
photometer
Optics in modular construction on compact cast base plate for stability and
endurance;
protection against moisture, exhaust fumes and chemical environmental impacts
Detector
Two-dimensional FFT backside illuminated CCD with high quantum efficiency and
increased UV sensitivity
Display modes
Emission
Counts (ct)
Intensity
Counts/seconds (ct/s)
Concentration
Value range: 5 characters (0.0001 to 99999), unit freely selectable
Signal evaluation
Spectrally resolved
Specters of 20 pixels to a maximum of 200 pixels in width
Power supply
Supply voltage
Frequency
230 V ±10%
50/60 Hz
Mains fuse installation in the
building
35 A
Typical average power
consumption
Base device: 4500 VA
Overvoltage category
II according to DIN EN 61010-1
Degree of contamination
2 according to DIN EN 61010-1
Protection class
I
Protection type
IP 20