![Analytical Technology Q45D/60 Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/analytical-technology/q45d-60/q45d-60_manual_2940581064.webp)
ATI
Q45D-ODO Optical D.O. System
Part 9– Troubleshooting
O & M Manual
Revision C (7/09)
63 -
MESSAGE
DESCRIPTION
POSSIBLE CORRECTION
D.O. Cal Fail
Failure of oxygen calibration. FAIL icon will
not extinguish until successful calibration
has been performed, or 30 minutes passes
with no keys being pressed.
Clean sensor redo zero and span calibration.
If still failure, sensor slope may be less than
20% or greater than 500%. Perform sensor
tests as described in section 10.4. Replace
sensor if still failure.
TC Cal Fail
Failure of temperature calibration. FAIL icon
will
not
extinguish
until
successful
calibration has been performed, or 30
minutes passes with no keys being
pressed.
Clean sensor, check cal solution temperature
and repeat sensor temp calibration. TC
calibration function only allows adjustments
of +/- 6 ºC. If still failure, perform sensor
tests as described in section 10.4. Replace
sensor if still failure. .
EPROM Fail
Internal nonvolatile memory failure
System failure, consult factory.
Chcksum Fail
Internal software storage error.
System failure, consult factory.
Display Fail
Internal display driver fail.
System failure, consult factory.
Range Cal Fail
Failure of factory temperature calibration.
Consult factory.
Figure 9-1 Q45D Display Messages (continued)
9.4
Sensor Tests
1. Check the condition of the optical sensing element. Mechanical damage to
the black covering over the optical element is an indication that sensor
problems are likely
.
2. Prior to disconnecting the sensor, measure the sensor output voltage at the
analyzer terminal strip with a DVM while the sensor is hanging in air. If the
sensor has been connected to a powered analyzer for at least 2 hours, the
nominal output of the sensor will be about +400mVDC when measured in air
at 25C (100% saturation) from BLACK (-) to RED (+) on the analyzer terminal
strips. This value is affected by temperature, pressure, and age of the sensor
so it’s possible to see a typical value that ranges from p200mVDC to
about +800 mVDC under a wide range of conditions in air.