Ax60K Carbon Dioxide Detector 
Quick Start Guide - Kiosk Option 

Document ref: P0159-804-01 

October 2015 

Page 7 of 8 

4

 

Operation 

4.1

 

Power on 

To power on the Ax60K, follow this procedure. 

[1]

 

Ensure the components are installed as described in section 3 Installation. 

[2]

 

Switch on the mains power supply at the wall socket. 
The Ax60K powers on and runs a 5-second self-test, during which: 
– the CO

2

 Alarm indicators illuminate 

– the CO

2

 Sensor indicators illuminate 

– the CO

2

 Sensor internal buzzer sounds 

Following a successful power-on, the CO

2

 Sensor begins continuously monitoring the air 

for CO

2

. During normal operation the status of the system is indicated as shown below: 

Normal 
operation 

with CO

2

 at a 

safe level 

CO

2

 Sensor indication: 

Power

 indicator flashes once per second. 

Alarm

 indicator is off. Buzzer is off. 

CO

2

 Alarm indication: 

Power

 indicator is on. 

Fault

 indicator is 

off. Strobe light is off. Sounder is off. 

5

 

Alarms 

 

WARNING: IF THE Ax60K DISPLAYS AN ALARM, YOU MUST FOLLOW THE SAFETY 

PROCEDURES SPECIFIED BY THE HAZARD WARNING AND INFORMATION LABELS. 

Alarms are indicated as follows: 

TWA alarm 
(0.5% 

average over 
the previous 

8 hours) 

CO

2

 Sensor indication: 

Power

 indicator flashes once per second. 

Alarm

 indicator flashes ¼ second on, 1¾ 

seconds off. Buzzer sounds in parallel. 

CO

2

 Alarm indication: 

Power

 indicator is on. 

Fault

 indicator is 

off. Strobe light is off. Sounder is off. 

Low alarm 

(1.5%) 

CO

2

 Sensor indication: 

Power

 indicator flashes once per second. 

Alarm

 indicator flashes 1 second on, 1 

second off. Buzzer sounds in parallel. 

CO

2

 Alarm indication: 

Power

 indicator is on. 

Fault

 indicator is 

off. Strobe light flashes 1 second on, 1 
second off. Sounder is off. 

High alarm 

(3%) 

CO

2

 Sensor indication: 

Power

 indicator flashes once per second. 

Alarm

 indicator flashes 

 second on, 

 

second off. Buzzer sounds in parallel. 

CO

2

 Alarm indication: 

Power

 indicator is on. 

Fault

 indicator is 

off. Strobe light and sounder are ½ 
second on, ½ second off. 

5.1

 

Acknowledging/clearing 

 

CAUTION: THE SENSOR RETAINS ITS CURRENT ALARM STATE, EVEN AFTER A 
POWER OUTAGE. IF AN ALARM IS NOT ACKNOWLEDGED AND THE Ax60K IS 

POWERED OFF, IT GOES STRAIGHT BACK INTO ALARM WHEN POWERED ON. 

To acknowledge/clear an alarm, follow this procedure: 

[1]

 

Press and hold 

Accept/Test

 until the buzzer sounds. The alarm is now muted. 

 

NOTE: THE ALARM IS NOW ACKNOWLEDGED. IT WILL AUTOMATICALLY 

CLEAR WHEN THE CO

2

 LEVEL REDUCES BELOW THE ALARM THRESHOLD. 

Содержание Ax60K

Страница 1: ...mited 15 Ellerbeck Court Stokesley Business Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net Ax60K Carbon Dioxide Detector Quick Start Guide Kio...

Страница 2: ...ANCE OF 3 METRES WARNING DO NOT TEST THE CO2 ALARM WHEN IT IS CLOSE TO THE EYES IT HAS A HIGH VISIBILITY STROBE LIGHT WITH A LUMINOUS INTENSITY OF 100 CANDELA WARNING ENSURE YOU PERFORM A RISK ASSESSM...

Страница 3: ...he level of carbon dioxide If the CO2 Sensor detects a level of CO2 above set limits it sends a signal to the CO2 Alarm The CO2 Sensor supplies power to the CO2 Alarm Power supply unit The power suppl...

Страница 4: ...with blue RJ45 connector 1 x mains power supply unit PSU plug in type complete with interchangeable heads to suit the destination country 1 x PSU securing strip 1 x RJ45 coupler for connecting the cab...

Страница 5: ...he CO2 Sensor Strobe light The strobe light is a very bright visible alarm NOTE The strobe window can be supplied in white blue or red The strobe light has two flash patterns 1 second on 1 second off...

Страница 6: ...ALL ACCESS AND EGRESS POINTS AND BUSY AREAS Retain the clear protective film on the fascia until the installation is complete Using the supplied paper template mark out the wall fixing position for th...

Страница 7: ...larm 0 5 average over the previous 8 hours CO2 Sensor indication Power indicator flashes once per second Alarm indicator flashes second on 1 seconds off Buzzer sounds in parallel CO2 Alarm indication...

Страница 8: ...hat the CO2 Sensor has a fault Power off the CO2 Sensor unit and call a service engineer 6 3 Calibration The Ax60K CO2 sensor is factory calibrated and does not require periodic calibration adjustment...

Отзывы: