![Analog Devices LTC6952 Скачать руководство пользователя страница 3](http://html1.mh-extra.com/html/analog-devices/ltc6952/ltc6952_demo-manual_2940197003.webp)
3
Rev 0
DEMO MANUAL DC2609A
QUICK START PROCEDURE
Figure 3. LTC6952Wizard Screenshot
Download the LTC6952Wizard setup file.
Run the LTC6952Wizard setup file and follow the instruc-
tions given on the screen. The setup file will verify and/
or install Microsoft.NET and install the LTC6952Wizard.
Refer to the LTC6952Wizard Help menu for software
operation.
DC2609A CONFIGURATION
1. Connect J30, J32 and J33 to the appropriate power
supplies and apply power (see Figure 1 and the Typical
DC2609A Requirements and Characteristics table).
2. Connect the DC2026 to the DC2609A with the provided
ribbon cable.
3. Run the LTC6952Wizard application.
4. In LTC6952Wizard, click File
→
Load Settings and
select file “LTC6952 EZSync StandAlone.6952set”.
The DC2609A’s red STAT LED (D1) should illuminate.
5. Synchronize outputs by toggling the LTC6952 SSYNC
bit on then off from the LTC6952Wizard.
6. Connect desired output (OUT9, OUT8, OUT5, OUT4,
OUT1 or OUT0) to a test instrument or other demo
board to evaluate performance.
Be sure to power down or terminate any unused RF output
with 50Ω, or poor spurious performance may result.