UG-866
EV-RADAR-MMIC2 User Guide
Rev. 0 | Page 10 of 23
EVALUATION AND TEST PROCEDURES
To evaluate and test the performance of the
ADF5901
,
ADF5904
, and
ADF4159
, use the following procedure:
1.
Install the
EV-RADAR-MMIC2
evaluation software.
2.
Connect the EV-ADAR-D2S board to the
EV-RADAR-
MMIC2
board.
3.
Connect the SDP board to the
EV-RADAR-MMIC2
.
4.
Connect the dc blocked TXOUT2 connector (J9) to the
spectrum analyzer.
5.
Connect 24.026 GHz, −50 dBm signal dc blocked from the
signal source to RX4 connector (J6).
6.
Run the
EV-RADAR-MMIC
software.
7.
Select the SDP board and the
EV-RADAR-MMIC2
(
ADF5901
,
ADF5904
, and
ADF4159
) device in the
Select
Device and Connection
tab of the software window.
8.
The software is set with default settings for the
following setup:
a.
Lock the
ADF4159
to 12.0125 GHz (24.025 GHz on the
ADF5901
).
b.
Calibrate the
ADF5901
to 24.125 GHz.
9.
In the
ADF5904
Main Controls
tab, click
Initialize
to
power up the
ADF5904
.
10.
In the
ADF4159
Main Controls
tab, click
Write All
Registers
to power up the
ADF4159
.
11.
In the
ADF5901
Controls
tab, click
Initialize ADF5901
to
power up and calibrate the
ADF5901
. See Figure 7 for a
typical setup.
12.
Measure a 24.025 GHz signal on TXOUT2 (J9) on the
spectrum analyzer
13.
Measure a 1 MHz baseband output signal on the
O5 connector on the EV-ADAR-D2S board.
Figure 7. Typical Evaluation Setup
PC
SDP
ADAPTOR
BOARD
POWER
SUPPLY
+5V
OSCILLOSCOPE/
SPECTRUM
ANALYZER
POWER
SUPPLY
+9V
–9V
SIGNAL
SOURCE
DC
BLOCK
SPECTRUM
ANALYZER
DC
BLOCK
1341
0-
00
7
Содержание EV-RADAR-MMIC2
Страница 1: ...13410 001...
Страница 3: ......
Страница 4: ......
Страница 5: ...13410 002...
Страница 6: ...13410 003...
Страница 7: ...13410 004...
Страница 8: ...13410 005...
Страница 9: ...13410 006...
Страница 20: ...13410 024...
Страница 21: ...13410 025...
Страница 22: ......