Quick Reference Guide Button-ology

Adjust Pressure Setting:

  The user can adjust the pressure 

in the main screen by pressing the 

up arrow 

button to 

increase the vacuum pressure or the 

down arrow 

button 

to decrease the vacuum pressure. The pressure may be 
adjusted in 1 mm/hg increments. Or, if arrow is HELD down,  
the pressure will increase/decrease in 10 mm/hg increments, 
to advance to the prescribed pressure setting quickly.

The large number in the center of the LCD screen will only 
change if the user physically changes it. Adjustments should 
be based on the physician’s order. 

The small number on the right side of the LCD screen (above 
the bars showing suction strength) indicates the pressure in 
the wound. This number will naturally fluctuate above and 
below the target pressure setting due to the permeability 
in the drape. Drape permeability prevents moisture build-up 
under the drape and helps ensure that intact skin remains 
clean and dry. If the small number drops 40 mm/hg or more 
below the prescribed pressure setting too many times in a 
minute, the pump will display an alert indicating there is  
a leak in the system.

Lock/Unlock Pump:

  Press and hold the 

Lock/Unlock  

and

 Exit/Cancel 

buttons

                   

at the same time for  

5 seconds. This will either lock or unlock the pump. When 
unlocked, you will notice a 

green open lock

 icon         on the 

bottom left of the LCD screen, or when locked, a 

red closed 

lock

 icon       . A locked screen disables the user from making 

any changes to pump settings.

Reset Treatment Time:

  Press and HOLD the 

Menu/Select

 

button         , press and RELEASE the 

On/Off

 button         , and 

then press and RELEASE the 

Lock/Unlock

 button         . Finally, 

RELEASE the 

Menu/Select

 button         .

Clear Alert Log:

  User MUST be in the main screen.  

Press and HOLD the 

Menu/Select

 button         , press and 

RELEASE the 

Lock/Unlock

 button         , and then press and 

RELEASE the 

Exit/Cancel

 button         . Finally, RELEASE the  

Menu/Select

 button         .

Silence Alert:

  Press the 

Menu/Select

 button         .  

(Note: the alert will sound again after 5 minutes if problem  
is not resolved.)

Setting Variable Pressure:

  Press the 

Menu/Select

  

button. The 

Treatment Mode

 may already be highlighted.  

If not, arrow up or down to highlight 

Treatment Mode

Once highlighted, press the 

Menu/Select 

button         . 

Arrow down to highlight 

Variable

. Once highlighted,  

press the 

Menu/Select

 button         ; this allows the user  

to access the 

Variable

 screen to adjust high/low time and 

high/low pressure settings.

Set Variable High Pressure Time:

  Press the 

up or down 

arrow

 button in 1 minute increments. This can be  

set from 1 to 15 minutes. To lock in time, press the 

Menu/Select

 button         .

Set Variable High Pressure:

  Press the 

up or down arrow

 

button to set high pressure. To lock in high pressure, 
press the 

Menu/Select

 button         .

Set Variable Low Pressure Time:

  Press the 

up or down 

arrow

 button in 1 minute increments. This can be  

set from 1 to 15 minutes. To lock in time, press the 

Menu/Select

 button         .

Set Variable Low Pressure:

  Press the 

up or down arrow

 

button to set low pressure. To lock in low pressure, 
press the 

Menu/Select

 button         .

Setting Intensity Mode:

  Press the 

Menu/Select

 button         . 

Use the 

up or down arrow

 button, to highlight 

Intensity

Once highlighted, press the 

Menu/Select

 button         . 

Press the 

up or down arrow

 button to set intensity to  

1, 2 or 3. The 3 settings of the intensity mode allow the 

user to adjust the suction speed of the XLR8+ device.  
Setting 1 will run at the lowest speed, while setting 3 will  
run at the highest speed. 

To lock in the intensity setting, press the 

Menu/Select

  

button         .

©2016 Genadyne Biotechnologies, Inc.   CLN-012   02/16

Questions?  

Call 1.800.208.2025 or 516.217.0010  |  email [email protected]  |  www.genadyne.com

XLR8/XLR8 Plus

Quick Reference Guide Button-ology

Adjust Pressure Setting:

  The user can adjust the pressure 

in the main screen by pressing the 

up arrow 

button to 

increase the vacuum pressure or the 

down arrow 

button 

to decrease the vacuum pressure. The pressure may be 
adjusted in 1 mm/hg increments. Or, if arrow is HELD down,  
the pressure will increase/decrease in 10 mm/hg increments, 
to advance to the prescribed pressure setting quickly.

The large number in the center of the LCD screen will only 
change if the user physically changes it. Adjustments should 
be based on the physician’s order. 

The small number on the right side of the LCD screen (above 
the bars showing suction strength) indicates the pressure in 
the wound. This number will naturally fluctuate above and 
below the target pressure setting due to the permeability 
in the drape. Drape permeability prevents moisture build-up 
under the drape and helps ensure that intact skin remains 
clean and dry. If the small number drops 40 mm/hg or more 
below the prescribed pressure setting too many times in a 
minute, the pump will display an alert indicating there is  
a leak in the system.

Lock/Unlock Pump:

  Press and hold the 

Lock/Unlock  

and

 Exit/Cancel 

buttons

                   

at the same time for  

5 seconds. This will either lock or unlock the pump. When 
unlocked, you will notice a 

green open lock

 icon         on the 

bottom left of the LCD screen, or when locked, a 

red closed 

lock

 icon       . A locked screen disables the user from making 

any changes to pump settings.

Reset Treatment Time:

  Press and HOLD the 

Menu/Select

 

button         , press and RELEASE the 

On/Off

 button         , and 

then press and RELEASE the 

Lock/Unlock

 button         . Finally, 

RELEASE the 

Menu/Select

 button         .

Clear Alert Log:

  User MUST be in the main screen.  

Press and HOLD the 

Menu/Select

 button         , press and 

RELEASE the 

Lock/Unlock

 button         , and then press and 

RELEASE the 

Exit/Cancel

 button         . Finally, RELEASE the  

Menu/Select

 button         .

Silence Alert:

  Press the 

Menu/Select

 button         .  

(Note: the alert will sound again after 5 minutes if problem  
is not resolved.)

Setting Variable Pressure:

  Press the 

Menu/Select

  

button. The 

Treatment Mode

 may already be highlighted.  

If not, arrow up or down to highlight 

Treatment Mode

Once highlighted, press the 

Menu/Select 

button         . 

Arrow down to highlight 

Variable

. Once highlighted,  

press the 

Menu/Select

 button         ; this allows the user  

to access the 

Variable

 screen to adjust high/low time and 

high/low pressure settings.

Set Variable High Pressure Time:

  Press the 

up or down 

arrow

 button in 1 minute increments. This can be  

set from 1 to 15 minutes. To lock in time, press the 

Menu/Select

 button         .

Set Variable High Pressure:

  Press the 

up or down arrow

 

button to set high pressure. To lock in high pressure, 
press the 

Menu/Select

 button         .

Set Variable Low Pressure Time:

  Press the 

up or down 

arrow

 button in 1 minute increments. This can be  

set from 1 to 15 minutes. To lock in time, press the 

Menu/Select

 button         .

Set Variable Low Pressure:

  Press the 

up or down arrow

 

button to set low pressure. To lock in low pressure, 
press the 

Menu/Select

 button         .

Setting Intensity Mode:

  Press the 

Menu/Select

 button         . 

Use the 

up or down arrow

 button, to highlight 

Intensity

Once highlighted, press the 

Menu/Select

 button         . 

Press the 

up or down arrow

 button to set intensity to  

1, 2 or 3. The 3 settings of the intensity mode allow the 

user to adjust the suction speed of the XLR8+ device.  
Setting 1 will run at the lowest speed, while setting 3 will  
run at the highest speed. 

To lock in the intensity setting, press the 

Menu/Select

  

button         .

©2016 Genadyne Biotechnologies, Inc.   CLN-012   02/16

Questions?  

Call 1.800.208.2025 or 516.217.0010  |  email [email protected]  |  www.genadyne.com

XLR8/XLR8 Plus

Quick Reference Button Guide