Jejunal (Glow Green
™
)
Feed Set
12
*
Attaching the Feed Set to the Jejunal Port:
1.
Remove jejunal feeding port safety plug (see page 7, #6) from the top of the
G-JET
®
Button, revealing the
Glow Green
™
color interlock.
2.
Holding the button, as shown in Figure A, line up the black mark on the jejunal
(
Glow Green
™
) feeding set adapter with the black mark on the G-JET
®
Button.
3.
Make sure the feeding set adapter is pushed in completely and gently turn
no more than 3/4 clockwise to lock the adapter in place. When fully locked, the
adapter will stop rotating.
Do NOT attempt to rotate the adapter past 3/4 turn.
If the adapter does not rotate
freely when first inserted into the feeding port, make sure the adapter is
pushed in
all the way
and that there is nothing blocking the jejunal port.
Remove the feed set by rotating the connector
COUNTER-CLOCKWISE
until the
black line of the feeding set aligns with the black line on the jejunal port. Remove the
feeding set and cap the jejunal port with the attached feeding port safety plug.
Starting Position
Fig. A.
Locked
Position
Applied Medical Technology, Inc.
Gastric (White)
Extension Set
13
*
Attaching Extension Set to the Gastric Port
1.
Remove the gastric decompression/medication port safety plug (see page 7, #7)
from the top of the G-JET
®
Button, revealing the white color interlock.
2.
Holding the button, as shown in Figure B, line up the black mark on the gastric
Decompression/Medication Extension Set with the black mark on the G-JET
®
.
3.
Make sure the extension set adapter is pushed in completely and gently turn the
adapter no more than 3/4 turn clockwise to lock the adapter in place. When fully
locked, the adapter will stop rotating.
Do NOT attempt to rotate the adapter past 3/4 turn.
If the adapter does not rotate
freely when first inserted into the feeding port, make sure the adapter is
pushed in
all the way
and that there is nothing blocking the gastric port.
Remove the extension set by rotating the connector
COUNTER-CLOCKWISE
until
the black line of the extension set aligns with the black line on the gastric port.
Remove the set and cap the gastric port with the attached port safety plug.
Starting Position
Fig. B.
Locked Position