background image

SE300 Series

 

Troubleshooting

 

 

18 

 

 

 

Common Issues (Optical)

 

 

Symptom

 

Cause

 

Solution

 

OPTICAL ISSUES

 

Obstructions  are 
observed in the field 
of view

 

Stains, dust, or dirt has accumulated on the speci- 
men

 

Clean the slide or use a new specimen if sample is 
destroyed

 

Stains, dust, or dirt have accumulated on the 
lenses

 

Clean the objective and eyepiece lenses

 

Unclear Image

 

Stain or dust has accumulated on the lens in the 
inlet of the head

 

Clean the lens  with lens cleaner or a nonabrasive 
lint free cloth, as well as spray with compressed air

 

The microscope head is not in the right position 
to be focused

 

Adjust the height of the microscope with the focus- 
ing  knob  on  the  focusing  rack  until  image  appears 
in  focus. 

Note: 

Higher  magnification  will  require 

the  head  to  move  closer,  while  lower  will  require 
more distance between the sample and the lens.

 

One side of the field 
of view is dark or the 
image moves while 
focusing

 

The specimen is not fixed

 

Secure  the  slide to  the  stage  with  clips  or  another 
manner so that the sample does not move

 

The field of view is not 
bright enough

 

The light source is not present

 

A light source must be used with this model micro- 
scope to view a sample (ring light, built in illumina- 
tor, or gooseneck illuminator)

 

The light source is too dim

 

The light source used is not bright enough, or is set 
to too low of an intensity setting. Adjust setting to a 
higher or change light sources

 

Stains, dust, or dirt has accumulated on the con- 
denser, objective, eyepieces, or base lens

 

Thoroughly clean tall lenses with lens cleaner or 
a lint free nonabrasive cloth

 

Содержание SE300 Series

Страница 1: ...SE300 Series 1 User s Manual ...

Страница 2: ...5 A SE306 A 9 Definition of Parts 10 Getting Started 11 Assembly 7 Operation 12 Setting Up 12 Focusing 12 Adjusting the Magnification 12 Attaching a Camera 13 Changing the Light Bulb 13 Maintenance Precautions 14 Specifications 15 Specifications 15 Electrical System 17 Objective Lens Parameters 17 Troubleshooting 18 Common Issues Optical 18 Common Issues Mechanical 19 ...

Страница 3: ...self with the features and functions of your new mi croscope If you d like more information on microscopes parts or accessories please visit our website at www iScopeCorp com We highly recommend you study this manual thoroughly before operating the microscope and that you keep it on hand for future reference If you have additional questions or need assistance please send us an email at info amscop...

Страница 4: ...ht or in high heat Keep it indoors in a dry and clean place with temperatures between 32 100 degrees F 0 40 degrees C and in maximum relative humidity of 85 3 Avoid touching the lenses on the objectives and the eyepieces so that oil and dirt from your fingerprints do not obstruct your view 4 Before turning the power on make sure that the power supply voltage is consistent with the voltage of your ...

Страница 5: ...SE300 Series Parts 5 SE303 P SE304 P Fig 1 SE303 P SE304 P Eyeguards Eyetubes Diopter Pillar Stand Lock screw Stage Clips Focusing Knob Top Light Objective Housing Stage Plate ...

Страница 6: ...SE300 Series Parts 6 SE305 P SE306 P Fig 2 SE305 P SE306 P Top Light Bottom Light Glass Stage Plate Objective Housing Focusing Knob Lock screw Pillar Stand Diopter Eyetubes Eyeguards ...

Страница 7: ...SE300 Series Parts 7 SE305R P SE306R P Fig 3 SE305R P SE306R P Lock screw Focusing Knob Reversed Head Top Light Bottom Light ...

Страница 8: ...SE300 Series Parts 8 SE307 P SE308 P Fig 4 SE307 P SE308 P Top Light Focus Knob Pillar Stand Bottom Light ...

Страница 9: ...SE300 Series Parts 9 SE305 A SE306 A Fig 5 SE305 A SE306 A Arm Stand Top Light Focusing Knob Bottom Light ...

Страница 10: ...e objectives are pre set atcertain levels and changed by rotating the objective FocusingKnob Used to move the microscope head vertically to bring the sample into focus Magnification Knob Changes the magnification of the lenses on a continuous zoom microscope Microscope Head Contains the objective lens eyepiece ocular tubes and prisms used tomagnify Ocular Tube Tube designed to hold the eyepieces o...

Страница 11: ... ve purchased and is generally listed on www amscope com at the items details page for your specific model 3 Remove the microscope body from the box and remove the plastic protective covering 4 Ensure that the item is fully intact and that the eyepieces are attached on top They are the black tubes that you look through the microscope to see your sample 5 Remove the cap from the objective lens if o...

Страница 12: ... the objective lenses Only models with DK come with these accessories Focusing 1 Adjust the focusing rack by loosening the lock screw located behind the pillar stand Move the entire focusing rack so that the image is roughly in focus then re tighten the lock screw If your model does not have a lock screw simply move on to the next step 2 Begin focusing by first looking with one eye through the eye...

Страница 13: ...dapter Note C mount cameras are not compatible with the SE300 Changing the Light Bulb 1 Turn off the power and unplug the microscope Wait until the bulb cools down 2 If change the top light s bulb loosen the lock screw on the lamp housing Take off the lamp cover remove the old bulb and replace with a new one Replace the lamp cover and re tighten its lock screw 3 If changing the bottom light s bulb...

Страница 14: ... lens cleaner we recommend Sparkle brand optical cleaner Do not use Sparkle to clean other elements of the microscope Use a neutral detergent on any plastic or painted surfaces Do not assemble or disassemble the microscope s electrical components yourself without advisement from one of our technicians Doing so will void your warranty unless by advisement of one of our technicians to do so After us...

Страница 15: ...SE300 Series Specifications 15 SE300 Series Specifications ...

Страница 16: ...SE300 Series Specifications 16 SE300 Series Specifications continued ...

Страница 17: ...40V 10 50Hz This electrical system is CE and GS certified 2 100V 120Vpowersupply 100V 120V 10 60Hz Thiselectrical system isULcertified All units come standard as 110V units unless an upgrade to a 220V system is requested Upgrade fee is dependent on which unit is purchased Objective Lens Parameters Magnification Field of View mm Working Distance mm 1x 20 53 2x 10 53 3x 6 7 53 4x 5 0 53 ...

Страница 18: ...age appears in focus Note Higher magnification will require the head to move closer while lower will require more distance between the sample and the lens One side of the field of view is dark or the image moves while focusing The specimen is not fixed Secure the slide to the stage with clips or another manner so that the sample does not move The field of view is not bright enough The light source...

Страница 19: ...de and utilize the stage clips No gem clip used Try using the gem clip for small items The objective touches the specimen Specimen is too large Get a smaller specimen Focusing rack too low Raise the focusing rack Head declines by itself The lock screw is too loose Tighten it firmly Focus knob does not turn Fine focusing limit reached Adjust the focusing rack s position w lock screw ...

Отзывы: