5-16
P129394-176
Operator Manual
Component Identification
5.11 Emergency Stop
Switch
An emergency stop switch (
Figure
) is a safety feature designed
to shut the sterilizer down completely in an emergency situation.
Pressing the emergency stop switch disconnects power to the door
and valves, causing the door to stop and most valves to close,
except S1 and S40, which are normally open, and will open when the
emergency stop switch
is pressed.
Important:
The emergency stop switch is for emergency use only!
DO NOT USE AS A START/STOP SWITCH
. To properly abort a
cycle, press abort button on sterilizer display.
Figure 5-12. Emergency Stop Switch, Key and Guard
Emergency Stop Label
Emergency Stop Switch
Emergency Stop Key
Emergency Stop Guard
Figure 5-13. Press Abort Button to End
Cycles When Needed
STATUS . . CHARGE
4
TEMP . . . 261.0 F
PRESS . . 27.7 psig
3.TRAY, PREVAC, 270F, 4, 30
PROJECTED CYCLE COMPLETION TIME:
12:20
MINUTES
SECONDS
PAPER
STATUS
FEED
ABORT
Abort
Button
–
Press when
necessary to
end cycle
Содержание C Series