Figure 11:
Optical Signal Conditioning
Each of the blocks depicted on Figure 11 can be individually enabled or disabled/bypassed.
4.2
ECG Amplifier
The ECG (electro cardiogram) amplifier is a high impedance, low noise instrumentation amplifier with
analog circuitry to band pass filter the signal. Gain is distributed between 3 gain stages. The gain in
the first stage determines the tradeoff between achievable noise level and achievable input offset
voltage. With the highest gain of 4 at the first gain stage (G1) about 400 mV of offset can be managed.
This value scales up to a max of 1.6 V of offset at gain 1. An optional 50/60 Hz notch filter can be
enabled to attenuate unwanted noise from mains coupling.
The recommended gain settings are 4-6-8 and 4-6-16