The following instructions are intended as a guide for
the surgeon . Various surgical techniques can be used
to implant the AMS penile prosthesis . The instructions
here represent one of these techniques .
CAUTION: This device is to be used only by
physicians who are knowledgeable regarding the use
of inflatable penile prostheses. This manual is not
intended to be a complete reference.
p
reoperatIve
s
etup
Instruments
The hospital should provide those instruments normally
required for a urological surgical procedure .
In addition to the AMS 700 Penile Prosthesis
components, you will need the following sterile setup:
9
Sterile normal saline (filling and flushing solution)
9
Two 60 cc and Two 10 cc syringes (for filling
and flushing prosthesis components)
9
Eight mosquito hemostats (for clamping tubing
when prepared with shods)
9
One pair of clean, sharp scissors for trimming
tubing
9
Hegar dilators (7 mm-14 mm), or urethral sounds
(21Fr-42Fr) (for dilating corpora cavernosa)
9
Furlow Insertion Tool (for measuring and
passing pulling sutures through glans)
9
AMS Quick Connect Assembly Tool (only
needed for sutureless window connectors)
9
AMS 700 with MS Pump Accessory Kit (see
description that follows)
9
AMS 700 with MS Pump Rear Tip Extender Kit
9
Cavernotomes (optional)
9
AMS Tubing Passers (optional)
9
AMS Closing Tool (optional)
9
SKW Retractor System (optional)
The AMS 700 with MS Pump Accessory Kit for the AMS
700 with MS Pump Product Line contains the materials
necessary for one implant procedure . It includes:
Special Purpose Needles
9
Two 15-gauge disposable blunt needles (for filling
components)
9
Two 22-gauge disposable blunt needles (for
flushing air and blood from tubing immediately
before a connection is made)
9
One pair Keith Needles (for transporting cylinder
pulling sutures through glans)
Note: The Keith needles are “lightening bolt” shaped -
the
bend is normal.
Hemostat Shods
9
Four 13 cm lengths of tubing (for covering tips of
hemostats used to prepare components—tubing-
shod hemostats help protect prosthesis from tubing
damage)
Tubing Connection Accessories
9
Four straight AMS Quick Connect Sutureless
Window Connectors
9
Three right angle AMS Quick Connect
Sutureless Window Connectors
9
One locking ring holder with eight collets
9
Three straight suture-tie connectors
9
Two right angle suture-tie connectors
9
One tubing plug (to prevent fluid from entering
or leaving the prosthesis during revision surgeries)
Documentation
9
One Quick Connect instructions for use brochure
9
One Patient Information Form (PIF)
9
One mailing envelope (for returning the
completed PIF to AMS)
9
One patient ID card
AMS Proximal Tool
The AMS Quick Connect Assembly Tool must
be ordered separately . It is a reusable stainless steel
instrument used to assemble the connectors .
The AMS Quick Connect system may be used for new
systems or when all previously implanted components
are removed and replaced with new components .
Operating Room Instructions
Содержание 700 CX
Страница 1: ...AMS 700 with MS Pump Penile Prosthesis Operating Room Manual English...
Страница 2: ...This page intentionally left blank...
Страница 4: ...This page intentionally left blank...
Страница 33: ...33 Appendix continued indications contraindications warnings precautions and potential adverse events Rx Only...
Страница 34: ...34 This page intentionally left blank...
Страница 35: ...35 This page intentionally left blank...