19. Perform Power On - Battery Check
20. Plug 2810E-2001 into AC power and allow battery to charge for 4 hours
(depending of the age of the battery and the surrounding temperature).
5.2.2
RECOMMENDED MAINTENANCE SCHEDULE
The following sections outline the recommended scheduled maintenance for the
2810E-2001.
1.
DAILY MAINTENANCE - Wipe off any accumulated dust or dirt on the front
panel or connectors using a clean, damp cloth. Pay particular attention to
where the various front panel components attach to the panel. Inspect the
outer case for any damage.
2.
WEEKLY MAINTENANCE - Wipe off any accumulated dust or dirt on the front
panel or connectors using a clean, damp cloth. Pay particular attention to
where the various front panel components attach to the panel.
3.
6 MONTH CHECK - Wipe off any accumulated dust or dirt on the front panel
or connectors using a clean, damp cloth. Pay particular attention to where
the various front panel components attach to the panel. Inspect the outer
case for any damage.
Inspect the PUSH-TO-TALK BUTTON, toggle switches and volume
controls for smooth operation.
Perform the 2-Wire and Full Duplex (4-Wire) system checks as described
in section 5.1.
5.2.3
YEARLY CHECK
For maximum service life, it is recommended that the chamber communicator be
sent back to Amron for a yearly check.
5.2.4
LONG TERM STORAGE
If the chamber communicator is to be stored for a period greater than 30 days, it is
recommended that it be stored in a cool dry location. Make sure that the POWER
SWITCH is turned off during storage. The 2810E-2001 should be stored plugged
into an AC outlet to ensure that the communicator will be fully charged and ready
to use when needed.