620UK-B Operation and Maintenance Manual – Rev J – January 2020 

Page 8

   

 

 

 

 

 

 

 

 

 

 

 

 

13231 Rooster Springs Rd. 

Austin, TX 78737 

T:

 (512) 858-4045 

E:

 [email protected] 

amptec.com 

and  “pin G” of the test lead connector (see 

diagram on right). The unit’s chassis ground can 

also be plugged into an earth ground connection 

(for example, at the test site to preclude any static 

voltages from building up, if desired). To prevent 

damaging the gold pins and sockets, only use 

AMPTEC test leads that are model 620UK-B 

compatible. When using AMPTEC test leads or 

probes, the notched connector end plugs directly 

into the single access notched main connector. 

Turn the connector collar clockwise until you feel 

the click of the mating connector. 

 

D-2.2 Alternate Connections and 4-Wire Kelvin Information 

The AMPTEC 620UK-B  meter is 

compatible with most AMPTEC 630 

series accessories. 620UK-B 

compatible test lead/probes use a 

Trident connector on the front panel as 

the main connection for lead/probe 

sets. The diagram on the right also 

provides insight on  how  the  620UK-B 

operates. The Trident connection 

provides the same pin socket arrangement, 4-Wire current high/low, and voltage sense 

high/low routed to the end of the leads, and the 4-Wire Kelvin wires then terminate in a 

variety of configurations depending upon the accessory. The 620UK-B  keyed access 

connector also makes it virtually impossible during normal operation to incorrectly connect 

lead/probe sets. 

 

The 620UK-B meter measures DC resistance with a 4-Wire Kelvin, or 4 terminal ohms, 

technique. Fundamentally, the 4-Wire Kelvin resistance method avoids measurement 

errors induced from the resistance of the test lead/probes and the contact resistance of 

the lead/probes to the device under test. Two of the wires, current high and current low, 

pass through the bond resistance under test in a current loop as part of a constant 

current source of 100 mA. It doesn’t matter if the current loop is longer or if there is 

some contact resistance with the connection under test, because the constant current 

source’s compliance voltage works to maintain 100 mA flowing in the loop. The other 2 

wires, voltage high and voltage low, sense the voltage drop across the resistance under 

test. The voltage measurement is high impedance. Meaning, virtually no current is 

drawn in either voltage sense lead wire. 

 

In conclusion, with 4-Wire Kelvin measurements, long distance test leads do not create 

a measurement error as long as you have enough compliance voltage remaining in the 

constant current loop. If test resistance is too high for the range, the meter will flash  

“1 - - - -” and “>>>OR” to indicate the meter is over range. 

 

Содержание 620UK-B

Страница 1: ...13231 Rooster Springs Rd Austin TX 78737 T 512 858 4045 E info amptec com amptec com AMPTEC RESEARCH MODEL 620UK B OPERATION AND MAINTENANCE MANUAL...

Страница 2: ...Feature Information p 9 D 3 620UK B Padded Carrying Case Photo and Information p 11 D 4 Detailed Use Description p 11 SECTION E MAINTENANCE REPAIR AND CALIBRATION E 1 General Maintenance Repair Calib...

Страница 3: ...on electrical subsystems used in general purpose aircraft helicopters fighter jets UAVs drones missiles etc Continuity determinations on fuses electrical cable connections heavy duty metal to metal c...

Страница 4: ...t has a low battery the display will say LO BAT To see instructions on battery replacement see section E 9 1 on page 17 4 Perform a visual test lead integrity check It is recommended that the user per...

Страница 5: ...is 20 000 counts 4 digit liquid crystal display with a back light that indicates autolock See section D 2 7 on page 9 for additional information about the autolock feature Low Battery Indication Displ...

Страница 6: ...d set with ground 630 305 4 Twin single banana jack lead set 630 401 4 Single point handheld Kelvin probes 630 405K 10 heavy duty spiked Kelvin probes 630 405S 10 heavy duty spear Kelvin probes 630 40...

Страница 7: ...tart updating the LCD screen once an electrical connection has been made After approximately 3 seconds the meter will produce a beeping sound and the backlight will turn on to illuminate the LCD scree...

Страница 8: ...wires then terminate in a variety of configurations depending upon the accessory The 620UK B keyed access connector also makes it virtually impossible during normal operation to incorrectly connect le...

Страница 9: ...id crystal display and high quality the display itself should have a life lasting many years Should the display become damaged or stop operating properly it can be replaced at an AMPTEC facility The u...

Страница 10: ...r to produce the final autolocked measurement which is the reading displayed while the backlight is on To assist the user the meter produces an audible beeping indicator along with the lit backlight o...

Страница 11: ...including front and rear panels Item C 4 Digit LCD Display with LO BAT and OR shown as an example Item D AUTO MEASURE RESET BUTTON Press and hold to reset a measurement that has been locked Item E Pow...

Страница 12: ...is on as the unit s autolock feature freezes the reading The backlit frozen measurement will stay frozen until you either press the AUTO MEASURE RESET button or press the test probes against a conduct...

Страница 13: ...uct lines and is capable of fixing nearly any issue your instrument may have AMPTEC maintains one sole facility for both the manufacturing and repair of its products AMPTEC RESEARCH CORP 13231 Rooster...

Страница 14: ...of test current On a 1 resistor the 620UK B 100mA output should get a 100mV drop 0 10V 0 10A For DVM scaling on the 2 range voltage drop is amplified using a x10 for range scaling hence a 10 mV drop b...

Страница 15: ...eck for a reading of 0 000 3 Put the output leads across a 0 100 resistance standard 4 The display should read 100 00 m 10 counts depending on the resistor s accuracy If the above test proves untrue t...

Страница 16: ...between readings for settling and an updated display of the measurement The meter will continue to sample and hold and then resample as long as it senses any continuity Finally connect the meter via t...

Страница 17: ...e the front panel and screws E 9 Battery Compartment Design and Description The metal battery compartment is powder coated to make it chemically resistant to battery leakage and backed with a water ti...

Страница 18: ...is certification was gained in 2013 and is maintained with quarterly inspections by an independent party AMPTEC also follows UL 913 s quality standards plant wide including the adoption of a quality m...

Страница 19: ..._________________________________________________________ ____________________________________________________________ ____________________________________________________________ ____________________...

Отзывы: