4
key once more to cancel the Back-Light function.
Back-Light on
J
lighter
J
Back-Light off
º»Ê ÕÌÌ\Ê`ÕÃÌÊÌ iÀVÕ«iÊvvÃiÌ
1. Insert the bead wire into a known temperature (T)
until the display equal to known temperature (T).
exp: ice point at 0°C boiling water at 100°C (Press
“
” key).
2. Press “
S
” or “
T
” to add or subtract the value.
3. It can be adjusted ±6°C (±10.8°F) of default. If you
V>½ÌÊ>`ÕÃÌÊÞÕÀÊ/É]Ê«i>ÃiÊV iVÊÞÕÀÊ/ÉÊÀÊÃi`Ê
the meter to be calibrated.
4. Press “
/,
” key to confirm.
º
S
É,»Ê ÕÌÌ
Press “
,
” key to enter Relative mode, zero the display,
and store the displayed reading as a reference value and
annunciator REL is displayed. Pressing “
,
” key over 2
seconds to exit the relative mode.
º
T
É"»Ê ÕÌÌ
Press the “
"
” key to enter the Data Hold mode,
the “
"
” annunciator is displayed at the higher
center of display. When HOLD mode is selected, the
thermometer held the present readings and stops all
further Measurements. Press the “
"
” key again
cancels HOLD mode, causing thermometer to resume
taking measurements.
º/£Ê/ÓÉ/£/Ó»Ê ÕÌÌ
Press “
/£
”, “
/Ó
”, “
/£/Ó
” key to select input mode T1,
T2 or T1-T2.
The input selection indicates which input is selected
for display. T1 thermocouple, T2 thermocouple or the
differential between the two thermocouples(T1-T2),
when the thermocouple is turned on, it is set to T1.