7
1
Press and hold down the switch to unlock.
2
Remove the antenna from the probe.
Removing the antenna in a manner not specified in this manual may cause
damages to the instrument and protection provided by the instrument may
be impaired.
Attaching and Removing the Adapter Plate
Figure 3: Attaching and
removing the adapter plate
Attaching adaptor plate
(clockwise)
Removing adaptor plate
(counterclockwise)
LAMP TEST
Non-contact voltage detection is always on. When the probe or the antenna
is close to energized wires, LED flashes and beeps. Non-contact voltage
detection will be disabled when LAMP TEST button is pressed.
W
• Verify lamp test function on a known lamp and/or on the identical
lamp model both before and after use to ensure the instrument is in
good working condition.
• On a very few particular linear fluorescent types, testing against the
glass surface of the lamp may not respond. Testing on the lamp socket
pin via the instrument’s probe (direct metal-to-metal contact) shall
allow good lamp response from a working lamp.
For better results, do the following while testing:
• Press and hold down LAMP TEST button for one second and release it
for one second.