7

Triangle, square, and sine waves

(1)  Select the 

Function (8) 

, select the 

Range (7)

, and rotate the 

FREQUENCY (13) 

knob to set the required 

frequency (read output in 

display window (4) 

).

(2)  Connect the 

Output (22) 

terminal to the oscilloscope for observing the waveform signal.

(3)  Rotate the 

AMPL (12) 

knob to control the waveform amplitude.

(4)  If attenuation of the output signal is required, pull out the 

AMPL (12) 

knob to obtain 20dB attenuation or press 

the 

ATT -20dB (12a) 

button for additional 20dB attenuation..

(5)  The phase-relation of Output waveform and TTL output is shown in Figure 5 below:

Figure 5

TTL PULSE   0V

TRIANGLE   0V

SINE   0V

SQUARE   0V

           

Pulse wave generation

(1) Press the 

Function (8)

 button ( 

 ), select the 

Range (7)

and rotate the 

FREQUENCY (13)

 knob to set the frequency.

(2) Connect the 

OUTPUT terminal (22)

 to the oscilloscope for 

observing the output signal.
(3) Pull out and rotate the 

Duty (9)

 knob to adjust the width of 

pulse waveform.
(4) Adjust the 

AMPL (12)

 knob to control the pulse amplitude.

(5) Pull out the 

AMPL (12)

 knob to attenuate the output signal by 20dB.

Ramp wave generation

(1) Press the 

Function (8)

 button ( 

 ) , select the 

Range (7)

 , rotate the 

FREQUENCY (13)

 knob to set the 

frequency.
(2) Connect the 

OUTPUT terminal (22)

 to the oscilloscope for observing the output signal.

(3) Pull out and rotate the 

Duty (9)

 knob to adjust the slope of ramp waveform.

(4) Adjust the 

AMPL (12)

 knob to control output amplitude of ramp waveform.

(5) Pull out the 

AMPL (12)

 knob to attenuate the output signal by 20dB.

TTL/CMOS signal output

(1) Select the 

Range (7)

 , and rotate the 

FREQUENCY (13)

 knob to set the frequency.

(2) Connect BNC connector of 

TTL/CMOS (20)

 to oscilloscope to observe the output signal.

(3) The output is a square waveform fixed at TTL level, suitable for general TTL integrated circuits.
(4) If a square waveform of CMOS level is required, pull out the 

CMOS (10)

 knob and adjust to required voltage 

levels.

Variation of external voltage-controlled frequency (VCF) (FG3C-UA only)

This mode of operation allows the user to adjust the frequency of the Function Generator with an external DC 
control Voltage. 
(1) Select the 

Function (8)

, select the 

Range (7)

, and rotate the 

FREQUENCY (13)

 knob to set the required frequency.

(2) Connect external control voltage (0 ± 10VDC) to the 

VCF (21)

 terminal via a suitable lead, and observe the signal 

generated from the 

Output (22)

 BNC terminal. 

(3) Other adjustments; the 

AMPL (12)

 knob can change the amplitude of signal, or attenuate the signal; the 

Offset 

(11)

 knob is for DC level changes, rotate the 

Duty (8)

 knob to change the output ratio of pulse or ramp waveform etc..

Sweep (internal) (FG3C-UA only)

(1) Select the 

Function (8)

 button and select the 

Range (7)

.

(2) Connect the 

OUTPUT (22)

 terminal to the oscilloscope  channel 1 for observing the OUTPUT signal. 

(3) Connect the rear panel 

GCV (23 )

 terminal to the oscilloscope channel 2 for observing sweep control signal.

(4) Rotate the

 FREQUENCY (13)

 knob to determine the upper output sweep frequency.

(5) Pull out the 

FREQUENCY (13)

 knob to activate sweep operation.

Содержание FG2C-UA

Страница 1: ...User Manual Mode d emploi Bedienungshandbuch Manuale d Uso Manual de uso FG2C UA FG3C UA Function Generator...

Страница 2: ......

Страница 3: ...FG2C UA FG3C UA Function Generator Users Manual Mode d emploi Bedienungshandbuch Manuale d Uso Manual de uso PN 2099447 April 2007 Rev 2 2007 Amprobe Test Tools All rights reserved Printed in China...

Страница 4: ......

Страница 5: ...tar libre de defectos de material y mano de obra durante 1 a o a partir de la fecha de adquisici n Esta garant a no cubre fusibles bater as descartables o da os que sean consecuencia de accidentes neg...

Страница 6: ......

Страница 7: ...from ground and never touch exposed wiring connections test probe tips or any live circuit conductors Do not operate instrument in an explosive atmosphere flammable gases fumes vapor dust Do not use t...

Страница 8: ...er and disposed of immediately as a plug with bared wires is hazardous if engaged in a live socket Any rewiring must be carried out in accordance with the information detailed on this label PREPARATIO...

Страница 9: ...P O W E R CM O S M IN Fig 1 FG2C UA Front Panel Fig 2 Rear Panel...

Страница 10: ...M k m H z P O W E R S W E E P WIDTH L I N A M F M D U T Y C M O S M I N M O D 1 M M O D Fig 3 FG3C UA Front Panel Fig 4 FG3C UA Rear Panel...

Страница 11: ...sh button on the panel as shown in Table 1 Table 1 Push Button Frequency Range 1 0 3Hz to 3Hz 10 3Hz to 30Hz 100 30Hz to 300Hz 1k 300Hz to 3kHz 10k 3kHz to 30kHz 100k 30kHz to 300kHz 1M 300kHz to 3MHz...

Страница 12: ...EXT Counter Switch FG3C UA Select internal counter mode count the frequency of the FG3C UA output or select EXT counter mode for an independent counter input signal from BNC 19 terminal 19 EXT Counte...

Страница 13: ...ramp waveform 5 Pull out the AMPL 12 knob to attenuate the output signal by 20dB TTL CMOS signal output 1 Select the Range 7 and rotate the FREQUENCY 13 knob to set the frequency 2 Connect BNC connec...

Страница 14: ...om 0 to 1 Vpp The FG3C UA output signal amplitude is set by the FG3C UA amplitude controls Note The sweep width can only be adjusted during the sweep cycle and it cannot be stopped AM FM Modulation in...

Страница 15: ...e rear of the Function Generator Warning Failure to turn off and remove power from the Function Generator before changing the mains power fuse could result in damage to the instrument and the operator...

Страница 16: ...check the Where to Buy section on www amprobe com for a list of distributors near you European Correspondence Address Amprobe Test Tools Europe P O Box 1186 5602 BD Eindhoven The Netherlands Correspon...

Страница 17: ...N A 0 to 5 Modulation Frequency N A 400Hz INT DC to 20 kHz EXT EXT Sensitivity N A 10Vpp for 10 modulation 12 Frequency Counter Int Ext N A Switch selector Range N A 0 3 Hz to 3 MHz 5 Hz to 150 MHz E...

Страница 18: ...CE Marking However electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement circuit Measuring instruments will also respond to unwanted signals...

Отзывы: