background image

Set-up procedure

Determine test loop location

Step 1.

 The test loop should be placed within the intended area to be looped and positioned at the   

             same height at which the final loop will be installed.

Step 2.

 Take note of the construction of the area to be looped and ensure the test loops cover all   

             the various structural elements. The test loops should cross over any steel beams or large   

             ducts that may be present. If the area to be looped is very large, for example a large   

             theatre, repeat the test in multiple areas throughout the space to account for variances in  

             building structure.

Step 3.

 Complete a separate site assessment procedure, or a metal loss test form for each test  

             area and note the location of each test. 

Perimeter Loop layout test loop

If a 

perimeter loop

 is going to be installed, then test with the actual intended loop layout.

Driver set-up

A.

 Connect the power supply (or battery), signal source (phone or MP3 player) and loop wire.

B.

 If using a multimeter this should be connected in series, using the high current connections 

and set to the AC current option.

C.

 Switch on the power to the driver and check that the power LED illuminates, the Input LED 

will also light up for a few seconds as the circuit settles. 

 

Signal level test set-up and test positions

Set the signal input to sinewave or combination and turn up the volume of the signal input until the 

input LED lights light up on the driver.
NOTE: continuous prolonged use of a sinewave signal (longer than 10 minutes) will cause the 

output current to drop slightly, however this should not affect the output level by more than 1dB. 

The combination signal avoids this.The current LED should also be on at this stage.

For 

perimeter loop 

tests, readings should be 

taken in the middle, edge and near to at least 

one of the corners of the loop, to measure the 

maximum variation in signal strength across the 

loop. 
 

All testing positions should be at least 1m (3ft) 

from the wires, as the signal level decreases near 

to the wire, making readings unreliable. 

Measurement height for the test meters should be at 1.2m (4ft) from the floor (typical seated head 

height), whether the test loop is on the floor or ceiling. 

All measurements must be carried out with the Loopworks Measure R1 or FSM positioned 

vertically.

HLS-2D Metal Loss Test Procedure

Page 3

Additional considerations:- The presence of fixed metal furniture and/or tiered seating may define 

the positions where a loop can be practically installed. If a perimeter loop cannot be used, the size 

of the figure of eight loops should be determined by the fixed features, e.g. loop widths should 

cover 2-3 steps with a 1 step gap.

Содержание HLS-2D

Страница 1: ...HLS 2D Metal Loss Test Procedure www ampetronic com...

Страница 2: ...and record sheet template available from www ampetronic com signals Optional equipment True RMS multimeter to monitor output current Additional short link cable 2x4mm connectors if multimeter is used...

Страница 3: ...will correct this There is no restriction on the width of the test loop below 15m 50ft and the designed loop width is likely to be dictated by installation practicalities e g aisle placement Technica...

Страница 4: ...it settles Signal level test set up and test positions Set the signal input to sinewave or combination and turn up the volume of the signal input until the input LED lights light up on the driver NOTE...

Страница 5: ...seconds as the circuit settles Signal level test set up and test positions Set the signal input to sinewave or combination and turn up the volume of the signal input until the input LED lights light u...

Страница 6: ...her follow the interpreting results section or send it to Ampetronic or your local distributor for analysis Define test points Start Test Start Site Assessment Test then name session Fill in test loop...

Страница 7: ...put signal to pink noise Step 5 Switch the meter mode to Frequency Response Step 6 Select the 100Hz band and record the signal level as shown on the white scale switch to 1kHz and repeat then to the 5...

Страница 8: ...op segment widths on the design match your test loop size your test result can be entered into the Use manual loss figure field on Inputs Design to specify the most suitable amplifier Interpreting fre...

Страница 9: ...ufactured to the following specifications LVD Safety Standard EN 60065 2014 Audio video and similar electronic apparatus Safety requirements A11 2017 RED Standard ETSI EN 303 348 Induction loop system...

Страница 10: ...www ampetronic com Ampetronic Ltd Tel 44 0 1636 610062 Email sales ampetronic com UP17940 3...

Отзывы: