ASS190 Slide Stainer
AMOSASS190SM20150814-01
---6---
4. The basket doesn’t run as program setting
The display board may be broken , please replace the display board as below picture .The program is
stored in display board .
5. The water tank don’t fill in with water .
It may caused by :
a.
Water supply interrupts . Solution: check
the water supply
b.
It is failure with the wiring of supply
electrovalve. Solution : Check the wiring
with multimeter and clear the trouble .
c.
The supply electrovalve is damaged .
Solution: replace a new one .
Display
board
Supply