Appendix A – Chapter 3 Mobile Patient Bed Maintenance
Part Number: 0160-6411, Rev A, 02/2019
Page 271 of 314
Software Release: cOS 5.00
Table 3.9 Wireless Quality of Service and Security
3.5 System Error Codes, Faults, and Messages
If the Mobile Patient Bed encounters an error or fault condition, laser treatment is interrupted;
the control pendant buttons blink; and/or an error message displays on the
CATALYS®
System
touchscreen.
NOTE
Refer to
Instructions for Interrupted Treatments
on page
for guidance on how to proceed
when treatment is interrupted.
Device
RF Wireless
Technology
Part Number
Operating
Frequencies
Operating
Distances
Industry
Standard
Quality of Service
and Security
CATALYS®
Mobile
Patient Bed
(MPB)
Bluetooth
Microchip
RN42
2.400 to
2.4835 GHz
Up to 10
meters
Bluetooth
Specification
version 2.1 +
Enhanced
Data Rate
(EDR)
FCC CFR47
Part 15,
Subpart C
“Intentional
Radiators”
The MPB wireless
communication has
low latency, high
data integrity, and
is robust in the
presence of
interference as
defined by the
Bluetooth
specification.
Security: The MPB
wireless
communication
channel is
configured to use
128 bit encryption.
Device System Wireless Functions, Capabilities, and Limitations
The
CATALYS®
Mobile Patient Bed (MPB) wirelessly and uniquely connects to a
CATALYS®
System
to allow the
CATALYS®
System the ability to enable and disable control features (joystick, pendant
buttons, etc.) and set motor speed limits based on the
CATALYS®
workflow.
Limitations during device operation: The MPB must be used within the specified operating dis-
tance of the connected
CATALYS®
System or the wireless connection will automatically disconnect
and the MPB will revert to its disconnected functionality.
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...