DESCRIPTION

The Amico Gas Control Panel is designed to deliver Nitrogen or Air to turbo-surgical tools. The Panel consists of a Shut-off 

valve, Delivery and Supply pressure Gauges, Pressure Regulator and DISS Outlet, in addition an outlet pipe is supplied 

that can be piped to additional Gas outlets.

RELIEVING and NON-RELIEVING

The panel comes in two styles “Relieving” and “Non-relieving”. A relieving panel allows gas to vent from the regulator, to 

reduce the pressure from the OUTLET PRESSURE side of the panel. CSA requires that a non-relieving panel be used where 

CSA Standards are practiced.

OPERATION

1.

 

Turn the Shut-off valve to the “Open” position.

2.

 

Verify that the correct supply pressure is available on the supply gauge.

3.

 

Attach a tool to the DISS outlet and tighten the nut (do not over tighten) or HP “Safety-Swing Coupling” outlet.

4.

 

Adjust the pressure regulator to the required pressure by observing the Outlet Gauge.

5.

 

Operate the surgical tool and observe the outlet pressure and adjust if needed.

6.

 

When the control panel is not in use reduce the outlet pressure to “zero” and turn off the supply valve.

INSTALLATION

Mount the back-box into the wall and adjust the adjustable side mounting brackets so that the top of the box will be 

flush with the finished drywall. Connect the inlet pipe to the gas pipeline and the outlet to additional outlets (this is 

optional). The pipeline can now be pressure tested to the required testing pressure. The outlet High Pressure Hose 

is shipped with one end connected to the outlet and the other end connected to the inlet on the nitrogen back box. This 

will enable the installer to pressure test the complete pipeline, including remote outlets, without having to attach the front 

assembly.

To attach the control panel front assembly to the control panel back box you must first disconnect the braided hose 

from the male PT adapter that is attached to the inlet pipe on the control panel back box. Then loosen and remove the 

panel cap from the male PT adapter that is marked “output” on the back of the control panel front assembly. Attach the 

braided hose from the back box to the male PT adapter (avoid over-tightening). Connect the “input” braided hose from the 

back of the control panel front assembly to the “input” of the male PT adapter in the back box (avoid over-tightening). 

Mount the control panel front assembly to the back box with the enclosed machine screws. 

PARTS LIST

Nitrogen control panel English DISS (Relieving type)    

 

N-CONP-E-REL

Nitrogen control panel English DISS (Non-Relieving type) 

 

N-CONP-E

Nitrogen control panel English “Dyna-Quip” HP (Relieving type)  

N-CONP-E-HP-REL

Nitrogen control panel English “Dyna-Quip” HP (Non-Relieving type)  N-CONP-E-HP

Nitrogen control panel French DISS (Non-Relieving type) 

 

N-CONP-F

Air control panel English DISS (Relieving type)   

 

 

N-CAIR-E-REL

Air control panel English DISS (Non-Relieving type) 

 

 

N-CAIR-E