Maintenance and Troubleshooting
| 5-27
Alarm Conditions and Corrective Action
This section lists all of the IPS-4 Analyzer system alarm messages (
Condi-
tion/Alarm Name
), describes what triggers the alarms, and provides action
that is required to reset the alarms (
Description and Suggested Corrective
Action
).
The system alarm relay contact opens on alarm. A system alarm can
result from any of the alarm conditions listed below. Some alarms may
not apply based on the system hardware configuration. For example,
systems without a heated sample system will not have a heater RTD
alarm.
Condition/Alarm Name
Description and Suggested Corrective Action
Heater Temperature
This alarm applies only to systems with a heated Oven/Sample System
Enclosure.
This alarm condition indicates the Oven Heater Plate temperature is
approaching its over-temperature limit; a soft shutdown of the heater
will occur. A soft shutdown means the system will shut down the heater
to ensure the temperature of the heater plate does not continue to
increase to ensure it will not exceed its temperature rating.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclosure
and the Oven/Sample System Enclosure doors, and:
•
Using an Ohm Meter, measure the resistance of the Oven Heater
Plate RTD and test it for an open circuit.
If the RTD is faulty, contact
AMETEK for assistance.
Cell Temperature
Bench Temperature
Oven Temperature
Note:
The Cell and Over Temperature alarms apply only to systems with a
heated Oven/Sample System Enclosure.
These alarm conditions indicate that the specific temperature-controlled
zone (Cell, Bench, Oven) is operating
below
or
above
its set point value
by
more
than 5 °C (typical; set point may vary).
If one of these alarms is caused by low temperatures in any of the
temperature-controlled zones, the analyzer will automatically switch to
Not Ready
mode to ensure the system does not become plugged due
to a low temperature in one of its zones.
This error can also be caused by a blown fuse, power interruption, or a
power spike, a failed heater, or wiring problems.
Corrective Action:
For alarms caused by
Low
/
High
Temperatures:
•
From the User Interface Panel:
- Check current operating temperatures for each temperature zone
and compare them to their set point temperatures.
- Check each of the associated configuration parameters for
temperature control. These parameters may have been lost as a
result of a power interruption or a power spike.
NOTE
Содержание Western Research IPS-4
Страница 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Страница 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Страница 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Страница 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Страница 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Страница 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Страница 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Страница 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Страница 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Страница 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Страница 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Страница 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Страница 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Страница 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Страница 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Страница 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...