Advanced Operation
DLM-E 3kW & 4kW Series Power Supplies
4.3.4
Remote Programming Only the Output Voltage or Current Limit
The front panel REMOTE/LOCAL switch allows you to switch back and forth between remote
and local programming signals for all three programming inputs of Voltage, Current, and OVP.
When operation is desired for programming only the output voltage and/or current limit without
the other, or to leave the OVP control on the front panel, follow the procedures below:
Remote Programming of the Output Voltage Only
For remote programming of output voltage only:
1. Turn off power to the supply.
2. Connect a programming source to remote programming connector
J3
between pins
9
(voltage programming input/positive) and
12
(return).
3. Connect a jumper between remote programming connector pins
J3-20
and
J3-21
for
external control of output voltage.
4. Adjust the external programming signal to zero.
5. Switch the
ENABLE/STANDBY
switch to
STANDBY
.
6. Turn the power supply
ON
.
7. Set the front panel
LOCAL/REMOTE
switch to the
LOCAL
position and adjust the
external programming signal. By pressing the
V&I PREVIEW
button and observing the
front panel voltmeter reading, the external control can be adjusted to the desired setting.
8. Adjust the current and
OVP
controls to the desired settings with the front panel controls
9. Switch
the
ENABLE/STANDBY
switch to
ENABLE
.
10. The output voltage is now remotely programmed with local control of the current limit
and
OVP
settings.
S1 Switch Settings
S1-1 OPEN = 0–5V
S1-1 CLOSED = 0–10V
Figure 4–8. Programming Output Voltage Remotely, Local Control of Current Limit/OVP
4-12
M362000-01 Rev E
Содержание Sorensen DLM-E
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...viii M362000 01 Rev E This page intentionally left blank...
Страница 8: ...x M362000 01 Rev E This page intentionally left blank...
Страница 10: ...xii M362000 01 Rev E This page intentionally left blank...
Страница 12: ...xiv M362000 01 Rev E This page intentionally left blank...
Страница 18: ......
Страница 44: ...Basic Operation DLM E 3kW 4kW Series Power Supplies 3 6 M362000 01 Rev E This page intentionally left blank...
Страница 70: ...Advanced Operation DLM E 3kW 4kW Series Power Supplies 4 26 M362000 01 Rev E This page intentionally left blank...