The
Measurement Settings
Menu sets the number of readings to average together
to reduce noise in the readback. Refer to Figure 3-53.
Figure 3-53. Measurement Settings Screen
The Measurement Settings menu has the following fields:
Entry
Description
Volt Avg Samples
Sets the number of voltage readings to average together to
reduce noise in the voltage readback. Allows to set a value
between 1 to 5. The value of 1 (factory default) provides the
fastest response time in the readings, but less rejection of noise.
Curr Avg Samples
Sets the number of current readings to average together to
reduce noise in the current readback. Allows to set a value
between 1 to 9. The value of 1 (factory default) provides the
fastest response time in the readings, but less rejection of noise.
3.1.7.6
A
UX
O
UTPUT
The
Auxiliary Output
Menu sets the 5V and 15V setting to
ON
or
OFF
state,
refer to Figure 3-54.
Figure 3-54. Auxiliary Settings Screen
The Auxiliary Settings menu has the following fields:
Содержание Sorensen Asterion AST Series
Страница 1: ...M330460 01 Rev A www powerandtest com DC Asterion 1U Operation Manual ...
Страница 2: ......
Страница 26: ......
Страница 53: ...DC Asterion Series Overview M330460 01 Rev A 2 27 This page intentionally left blank ...
Страница 54: ......
Страница 109: ...DC Asterion Series Overview M330460 01 Rev A 3 55 This page intentionally left blank ...
Страница 110: ......
Страница 112: ......
Страница 117: ...DC Asterion Series Index M330460 01 Rev A 3 ...
Страница 118: ...Maintenance DC Asterion Series 4 M330460 01 Rev A This page intentionally left blank ...