4.1.1  Laser Head Filter  

 
The laser filter can be set to improve signal stability via the Orbit Network. 
 
Available Settings:  20 kHz, 4 kHz, 1 kHz, 200 Hz, 25 Hz, 1 Hz and 0.1 Hz 
 
Note that on certain targets, particularly dark targets, setting the filter to 4 kHz or lower may improve performance as the CMOS 
integration time is increased at settings of 4 kHz or lower. 
When a filter other than 20 kHz is chosen, the effective sampling rate of the head will be reduced from 40 kHz to 20 kHz. This will permit a 
longer integration time on the CMOS sensor, which can be useful on absorbing (dark) targets. 
 
Important: Factory Default = 4 kHz. 
 

4.1.2  Level Cut Time 

 
Level Cut Time is used to modify the operation of the Laser head when the received laser intensity is below the dark cut value (~10%) or 
above the bright cut value (~75%). 
When level cut time is defined, instead of outputting the maximum / minimum displacement value (125% of head range), the Laser head 
will continue to output the value measured before the intensity went out of range. 
After the cut time expires, the maximum / minimum displacement value will be output. 
This function is generally used to handle situations when the target has known features that would cause the head to move to a saturated 
value. 
An example of this is a plastic plate that has holes drilled in it. When the laser beam moves into the holes, the value will go out of range. 
Using the cut time feature, the value before the laser went into the hole will be held. 
 
The unit is 0.1 msec. The minimum value is 0 msec. The maximum cut time is 1000.0 msec. 
Factory Default = No cut time. 
 

Содержание Solartron Metrology Orbit LTH

Страница 1: ...Page 1 of 15 503158 ISSUE 3 EDCR21694 ...

Страница 2: ...IPLES 9 3 1 SYSTEM DESCRIPTION 9 3 2 INSTALLATION AND SET UP 9 3 2 1 Environments that are known to cause problems 9 3 2 2 Head Orientation 10 3 2 2 1 Moving Targets 10 3 2 2 2 Slots and Walls 11 3 2 3 Mounting 12 3 3 LED INDICATOR DIFFUSE AND SPECULAR MODE 13 4 KEY FEATURES 14 4 1 1 Laser Head Filter 14 4 1 2 Level Cut Time 14 4 1 3 Laser Beam Control 15 5 Orbit and ORBIT ACS INTERFACES 15 ...

Страница 3: ...SUE 3 EDCR21694 FDA NOTICE This product complies with FDA CFR Title 21 Parts 1040 10 and 1040 11 excepting for variations relevant to FDA Laser Notice No 50 June 24 2007 IEC NOTICE This product complies with IEC 60825 1 2007 ...

Страница 4: ... Orbit ACS products offers a solid non contact metrology solution for many applications There are two versions of the laser the LTH Laser which is the highest performing laser in the range and the LTM laser which has lower resolution In this manual where information is common to both laser products the term LTX is used to mean either LTH or LTM ...

Страница 5: ...DCR21694 2 SAFETY WARNINGS AND CAUTIONS 2 1 WARNINGS AND CAUTIONS Symbol Definitions and Explanation Electrical Shock Electrocution Hazard Laser Aperture Class 3R Is considered safe if handled carefully with restricted beam viewing ...

Страница 6: ...cted beam may inadvertently be directed into the eye of the user WARNING DO NOT STARE DIRECTLY INTO THE BEAM This product emits laser radiation DO NOT stare directly into the beam with the naked eye or any form of optical equipment DO NOT stare at reflections of the beam DO NOT point the laser at a person s eye WARNING DO NOT USE TO PROTECT THE HUMAN BODY This device is not intended to be used in ...

Страница 7: ...ng flammable liquids flammable gases or flammable particulate matter WARNING DO OPEN OR MODIFY THE LASER HEAD Laser radiation levels inside the head can exceed the rated levels outside of the head Modification of the laser head may expose users to unsafe laser radiation levels beyond the rated levels of the head There are no user serviceable parts inside the laser head ...

Страница 8: ...cover will not be provided Laser ON LED This product has a range indicator LED which also functions as a Laser ON indicator When the range LED indicator is on any color the laser diode is activated 2 3 HAZARD AREAS Although the lasers are not considered hazardous during normal operation do not stare directly into the laser beam Due to the hazard potential associated with a class 3R and 3B lasers o...

Страница 9: ...LTX is compatible with all other Solartron Metrology Orbit products and can be used to form complex measurement systems combining both contact and non contact measurements 3 2 INSTALLATION AND SET UP The LTX operates by emitting a coherent light source and monitoring the reflected response Using this product in environments that impairs optical operations will adversely affect the performance 3 2 ...

Страница 10: ... ISSUE 3 EDCR21694 3 2 2 Head Orientation 3 2 2 1 Moving Targets The orientation of LTX relative to a moving target can affect the performance as a general rule position the wide face of the LTX perpendicular to the motion ...

Страница 11: ... 11 of 15 503158 ISSUE 3 EDCR21694 3 2 2 2 Slots and Walls When measuring near a wall or measuring in a slot ensure that the reflected beam is not blocked or reflected away from the head by any obstruction ...

Страница 12: ... surfaces which may be used to align the head for optimal performance The primary surface A is formed by the raised mounts on bottom of the head The second B is the front of the head and the third closest to the emitter When mounting the head the goal is to have the Reference Surface B parallel as possible to the plane of the target Take care when using mechanical blocks etc to set up the laser as...

Страница 13: ...Page 13 of 15 503158 ISSUE 3 EDCR21694 3 3 LED INDICATOR DIFFUSE AND SPECULAR MODE The green led indicates the most accurate position for measurement ...

Страница 14: ...sed to modify the operation of the Laser head when the received laser intensity is below the dark cut value 10 or above the bright cut value 75 When level cut time is defined instead of outputting the maximum minimum displacement value 125 of head range the Laser head will continue to output the value measured before the intensity went out of range After the cut time expires the maximum minimum di...

Страница 15: ...ORBIT ACS INTERFACES The LTX comes connected to either an Orbit module or an Orbit ACS module All of the laser functions can be set either using the Orbit Interface controllers such as the USB Interface Module and the Orbit Support Pack for Windows Orbit commands to set up the laser can be found in the Orbit Software Manual available on the Solartron Metrology Web Site Instructions for using the O...

Отзывы: