5.18 Calibration
radEAGLET
User Manual
Setting
Neutrons
Integration Duration
-5
4
4 sec
-4
4
10 sec
-3
4
15 sec
-2
4
20 sec
-1
4
30 sec
Default ±0
4
40 sec
+1
4
80 sec
+2
4
100 sec
+3
4
140 sec
+4
4
180 sec
+5
4
220 sec
+, -
Increase or decrease the value.
Accept
Make your change permanent.
Exit
Return to the superior screen.
Down
Highlight the next item. Cycles to the top item after you reached the last item.
Select
Select the highlighted item.
5.18 Calibration
This screen (Fig.
, p.
) shows a partial spectrum and some additional information you
can use to inspect and trigger the calibration of your
radEAGLET
.
For additional information refer to appendix
, p.
A simpler version of calibration is available too, see
, p.
Please place a suitable calibration source (
40
K or
137
Cs) in front of the detector before
proceeding.
Settings and Commands
Next
Reveal the next available command for the left key.
Source
Specify the source to be used for calibration. Available are
40
K and
137
Cs.
72/140
Software 3.2.0 • Document 3.2.2o
© innoRIID GmbH • 2018-10-24
Содержание ORTEC radEAGLET
Страница 18: ...18 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Страница 30: ...30 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Страница 38: ...38 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Страница 42: ...42 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Страница 90: ...90 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Страница 116: ...116 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Страница 132: ...132 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...