radEAGLET-R
User Manual
Using the
radEAGLET-R
For further processing screenshots, transfer them to your computer with, e.g., Storage
Management in the Web interface (see
, p.
), or a USB flash drive (see
p.
2.3 Starting Up the
radEAGLET-R
1. Position the instrument in a low-radiation environment.
If your model is not equipped with an internal source, position the supplied KCl (potta-
sium chloride) box (Fig.
, p.
) in front of the detector cap.
Figure 16:
Box with potassium chloride (KCl) for stabilization and calibration (Only
radEAGLET-R
models
without internal source (see
, p.
).)
2. Press and hold down the
key (Fig.
, p.
).
3. After a couple of seconds the instrument begins to boot when the alarm LEDs (Fig.
p.
) begin to flash.
4. You are welcomed by the start-up screen with dots appearing from left to right
(Fig.
, p.
5. The
radEAGLET-R
begins an initial stabilization (Fig.
, p.
, see
, p.
for de-
tails).
6. When the stabilization source is identified it is shown (Fig.
, p.
).
7. After the stabilization the
radEAGLET-R
switches to Dose Rate Mode (see
, p.
8. Watch the stabilization status icon on the screen (see
, p.
© innoRIID GmbH • 2020-11-18
Software 3.2.12r • Document 3.3.3b4o
31/163
Содержание Ortec radEAGLET-R 2SG
Страница 1: ...radEAGLET R User Manual Software 3 2 12r Document 3 3 3b4o 2020 11 18...
Страница 12: ...12 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Страница 22: ...22 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Страница 42: ...42 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Страница 94: ...94 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Страница 120: ...120 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Страница 143: ...A Nuclide Library to be defined innoRIID GmbH 2020 11 18 Software 3 2 12r Document 3 3 3b4o 143 163...
Страница 144: ...144 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...