background image

7

resolution of the null measurement. 

The  actual 

meas urement  

of 

linearity  is

accomplished  by  dialing the Pulse Height  dial  to  0,
resulting  in  the  amplifier  output  being  reduced  to
zero.  Since  the  Pulser supplies   s ignals  in  parallel
both  to  the  bridge  for  null  and  to  the  amplifier,
varying  the Pulser output will  have  no  effect  on  the
null if perfect amplifier linearity is assumed. 

As  an  ex ample  of  this  method,  assume  that  the
amplifier  under  test  has  essentially  zero  output
impedance. Set R1  equal to  100

Ù

  and  R2  equal  to

200

Ù

.  Let  diodes D1 and D2 be  2N2048  connected

as   diodes.  Only  one-half  of  the  actual  amplifier
output  voltage  can be measured  directly  at  point  A
due to the superposition of the outputs  of the Pulse
generator via R1 and the amplifier via R2. To specify
nonlinearity  as  a  percentage  of  full  Output  voltage,
the calibration of 10 mV/cm will be equal to 10 mV/5
V  or  0.2%  per cm. Therefore it is seen that 0.1%  is
quite easily resolved. 

In  addition  to  linearity  measurements,  it  is  obvious
that  this  method  can  be  quite  useful  in
measurements of temperature stability. 

Pulse  Height  Analyzer  Calibration 

  With  the

Pulser calibrated to read directly in terms  of  energy
as described earlier in this section, the calibration of
a complet e  spectrometry  system  from  preamplifier
to  multichannel  analyzer,  i.e.,  analog  to  digital
convert er  (ADC),  can  readily  be  accomplished  by
simply  feeding  into  the  preamplifier  a  calibrated
energy  signal  and  observing  the  corresponding
channel into which it is assigned by the ADC. 

An  important  consideration  in  this  test  involves
ensuring that the linear system "goes through zero,"
and that the output of the pulse generator is properly
terminated. The attenuator switches in the 480 have
an  accuracy  controlled by  1%  metal  film  resistors
and could be  used  to  digitally  check the linearity of
the  spectrometer.  In  addition  to  the  attenuator
accuracy, the Pulse Height control has independent
integral  nonlinearity  of  ±0.25%.  This  control
therefore  allows  an  integral  linearity  curve  of  the
ADC  to be  taken over the  continuous range  of  the
ADC, i.e., from zero to the maximum address of the
ADC.  Due  to  the  better  integral  linearity   control,
continuous scanning with the Pulse Height control is
the  recommended method  of  c hecking  for  system
linearity.  The linearity of  the  ADC  can  therefore  be
determined  by  having  previously  taken  the  linearity
curve  of  the  amplifier  and  preamplifier  as  outlined
earlier in this section. 

5.  MAINTENANCE

5.1.  TESTING PERFORMANCE OF THE

PULSER

 
The  following  information  is  intended  as  an  aid  in
the  ins t allation  and  checkout  of  the  480.  These
instructions  present  information  on  front  panel
controls ,   waveforms  at  test  points,  and  output
connectors. 

The  following,  or  equivalent,  test  equipment  is
needed: 

Tektronix Model 580 Series Oscilloscope 
100

Ù

 BNC Terminators 

Vacuum Tube Voltmeter 

Before testing the performance  of  the  480,  take the

following preliminary steps: 

1.  Visually  check  the  module  for  possible  damage
due to shipment. 

2.  Connect  ac  power  to  NIM-standard  Bin  and
Power Supply, 0RTEC 401 /402. 

3.  Plug  module  into  Bin  and  check  for  proper
mechanical alignment. 

4.  Switch  ac  power  on  and  c heck  the  dc  Power
Supply voltages at the test points on the 402. 

The performance test consists of the following: 

1. Set the front panel controls on the 480 as follows: 

a. relay switch to On, 

Содержание ORTEC 480

Страница 1: ...Model 480 Pulser Operating and Service Manual Printed in U S A ORTEC Part No 733390 0815 801 South Illinois Avenue Manual Revision C Oak Ridge Tennessee 37830 United Statesof America ...

Страница 2: ...o that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the fault of the instrument being returned and of the model serial and revision Rev on rear panel numbers Failure to do so may cause unnecessary delays in getting the unit repaired The ORTEC standard proc...

Страница 3: ... 2 2 4 ELECTRICAL AND MECHANICAL 2 3 INSTALLATION 2 3 1 GENERAL 2 3 2 CONNECTION TO POWER 2 4 OPERATION 3 4 1 PANEL CONTROLS 3 4 2 INITIAL TESTING AND OBSERVATION OF PULSE WAVEFORMS 3 4 3 CONNECTOR DATA 3 4 4 TYPICAL OPERATING CONSIDERATIONS 4 5 MAINTENANCE 7 5 1 TESTING PERFORMANCE OF THE PULSER 7 5 2 ADJUSTMENT OF DECAY TIME OF OUTPUT PULSE 8 5 3 TABULATED TEST POINT VOLTAGES 8 5 4 SUGGESTIONS F...

Страница 4: ...s a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please read all safety instructions carefully and make sure you understand them fully before attempting to use this product In addition the following symbol may appear on the product ATTENTION Refer to Manual...

Страница 5: ...the instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug the instrument from the ac power supply Remove loose dust on the outside of the instrument with a lint free cloth Remove remaining dirt with a lint free cloth dampened in a general purpose detergent ...

Страница 6: ...vi ...

Страница 7: ...is not possible to overload the Bin Power Supply with a full complement of modules in the Bin However this may not be true when the Bin contains modules of other than ORTEC design All signal levels and impedances are compatible with other modules in the ORTEC 400 Series 1 2 BASIC FUNCTION The 480 provides output pulses that are characterized by a fast rise time and a slow exponential decay time Th...

Страница 8: ...r negative dc coupled output with an impedance of 100Ù Direct Front panel BNC connector provides positive or negative dc coupled 0 to 10 V pulse into a high impedance and 0 to 5 V max pulse into 100Ù This is equivalent to a range of 0 to 220 MeV energy referred to a silicon detector when used with associated charge terminator Accessories Included One 100Ù voltage terminator and one charge terminat...

Страница 9: ...y to be driven from the ac line The frequency of the ac line will be 50 to 60 Hz Neg Pos The Polarity of the output signal will be either negative or Positive as determined by the setting of this front panel slide switch 4 2 INITIAL TESTING AND OBSERVATION OF PULSE WAVEFORMS See Section 6 1 for test performance data 4 3 CONNECTOR DATA CN 1 The Direct Output BNC connector provides a dc coupled outp...

Страница 10: ...divisions is equivalent to a specific MeV loss in a radiation detector The procedure is as follows 1 Connect the detector to be used to the spectrometer system i e preamplifier main amplifier and biased amplifier 2 Allow particles from a source of known energy alpha particles for example to fall on the detector 3 Adjust the amplifier gains and the bias level of the biased amplifier to give a suita...

Страница 11: ...llustrated in Fig 4 3 The amplifier noise resolution spread can be measured correctly with a pulse height analyzer and the 480 as follows 1 Select the energy of interest with the 480 and set the linear amplifier and biased amplifier gain and bias level controls so that the energy is in a convenient channel of the analyzer 2 Calibrate the analyzer in keV per channel using the purser Full scale on t...

Страница 12: ...st produce a positive Output Pulse The impedance seen from point A to ac or signal ground via point C should be equal to the impedance seen from point A to ac or signal ground via point B The diodes D should be germanium units with high gm The diodes can be replaced with high frequency germanium transistors with the base connected to the collector so that the emitter base functions as the diode Tr...

Страница 13: ...ves ensuring that the linear system goes through zero and that the output of the pulse generator is properly terminated The attenuator switches in the 480 have an accuracy controlled by 1 metal film resistors and could be used to digitally check the linearity of the spectrometer In addition to the attenuator accuracy the Pulse Height control has independent integral nonlinearity of 0 25 This contr...

Страница 14: ...switched in the output pulse should be between the following limits 9 Set the Polarity switch to Neg There should be no change in amplitude from the Pos position Observe the output with a sweep of 5 msec cm and look for skipping or other erratic behavior of the relay 5 2 ADJUSTMENT OF DECAY TIME OF OUTPUT PULSE As the 480 is normally supplied the decay time of the output pulse is essentially fixed...

Страница 15: ...ting instructions of Section 6 1 of this manual and the circuit description in Section 5 are intended to provide assistance in locating the region of trouble and repairing the malfunction The guide plate and shield cover can be completely removed from the module to enable oscilloscope and voltmeter observations with a minimum chance of accidentally short circuiting portions of the etched board The...

Страница 16: ...eserved bus 34 Power return ground 13 Spare 35 Reset Scaler 14 Spare 36 Gate 15 Reserved 37 Reset Auxiliary 16 12 V 38 Coaxial 17 12 V 39 Coaxial 18 Spare bus 40 Coaxial 19 Reserved bus 41 117 V ac neutral 20 Spare 42 High quality ground 21 Spare G Ground guide pin 22 Reserved Pins marked are installed and wired in ORTEC s 4001A and 4001C Modular System Bins Bin Module Connector Pin Assignments Fo...

Отзывы: