User Guide

Landcal Blackbody Source

Type R1500T

Page 5

4  Electrical Supply

4.1  Electrical connections

A suitable single phase a.c. electrical supply with earth (ground) will be 

necessary.  The supply may be live to neutral non reversible, live to neutral 

reversible or live to live.  The furnace must be connected to the electrical supply 

for which it was ordered.  Details of the voltage, current and frequency required 

are stated on the furnace plate attached to the furnace.  The supply point must 

be within reach of the operator and must incorporate either an isolating switch 

or a quickly removable plug.
If the furnace is supplied with a mains cable fitted then it may be wired directly to 

an isolator or fitted with a line plug.  Supply rated fuses will be internally fitted.
If the furnace is supplied without a mains cable, the specification for suitable 

cable for 220/240V operation is: 

300/500 volt, PVC insulated 3 core cable. Each core 4mm ² for 110/120V 

operation.
300/500 volt, PVC insulated 3 core cable. Each core 2.5mm ² for 220/240V 

operation.

The colour code for the cable is:

Brown lead 

 

 

Live

Blue lead   

 

 

Neutral

Green Yellow lead   

Earth (ground)

As no internal supply rated fuses are fitted, a permanent connection to an 

appropriately fused and isolated supply must be made to the internal terminals 

after temporary removal of the furnace panel.

4.2  Fuses

Refer to the circuit diagram, Fig. 2.
F1  Internal supply fuses are only fitted if a mains cable is fitted.  The supply to 

the furnace should be fused to the same rating.
110/120V  

 

 

40 amp fuse

220/240V  

 

 

25 amp fuse

F2  Instrument circuit fuses  32mm x 5mm glass, 2 amp
F3  Heat light fuses 

 

32mm x 5mm glass, 2 amp   

  

 

 

 

 

Not fitted for 220/240V operation 

  

 

 

 

 

Fitted for 110/120V operation

4.3  Earth leakage

Warning

It is recommended that this furnace is not used in conjunction with an 

earth leakage trip switch.  At elevated temperatures all ceramic materials 

become  slightly  conductive.    This  may  mean  that  at  temperatures 

above 1000°C a leakage current in excess of 30 milliamps is present.  

This will cause an earth leakage trip switch to activate.

Содержание LAND R1500T

Страница 1: ...le 01246 410585 Email land infrared ametek co uk Internet www landinst com Land Instruments International 2008 2012 AMETEK Land Inc 150 Freeport Road Pittsburgh PA 15238 U S A Telephone 412 826 4444 F...

Страница 2: ...at safety precautions are taken whenever the materials are handled Exposure to fibrous dust may cause respiratory disease When handling the material always use an approved respiratory protection equip...

Страница 3: ...n It is possible that this equipment is to be installed on or near to machinery or equipment operating at high temperatures and high pressures Suitable protective clothing along with face and eye prot...

Страница 4: ...s etc Design and Manufacturing Standards The Quality Management System of Land Instruments International is approved to BS EN ISO 9001 for the design manufacture and on site servicing of combustion en...

Страница 5: ...ription 3 3 Specification 4 4 Electrical Supply 5 5 Commissioning 7 6 Using the Furnace 12 7 Calibration of Radiation Thermometers 13 8 Maintenance 14 9 Eurotherm Temperature Controller Type 3216CC 16...

Страница 6: ......

Страница 7: ...ller indication From previous work the indicated temperature has been found to agree with the radiance temperature to within 10 C at 1000 C when using thermometers operating at short wavelengths 2 The...

Страница 8: ...some individuals be an irritant to the skin nose and throat Safety Note Refractory Fibrous Insulation This product contains alumino silicate wool products in its thermal insulation These materials ma...

Страница 9: ...lection of standards will depend upon customer requirements e g working temperature range type of radiation thermometers to be calibrated etc Ideally the standards should be of a similar type to those...

Страница 10: ...stalled in the source the distance from front face to cone point is 145mm 5 7in External aperture Approx 38mm 1 5in Emissivity Approximately 0 99 at short wavelengths Resistance heating elements 220V...

Страница 11: ...m for 110 120V operation 300 500 volt PVC insulated 3 core cable Each core 2 5mm for 220 240V operation The colour code for the cable is Brown lead Live Blue lead Neutral Green Yellow lead Earth groun...

Страница 12: ...ype R1500T CA970249 4 4 Wiring diagram GR Y L N E TC 2 3 5 9 Pin D connector RS 232 SSR ELEMENT S Mains filter L N R R Thermocouple HEAT PI G BU R BU Instrument switch Supply light F2 F2 F1 F1 F3 F3 R...

Страница 13: ...ng elements and a set of element support brackets braids are supplied separately These items should be fitted as follows 1 Remove the stainless steel front panel unscrewing the four equally spaced scr...

Страница 14: ...Page 8 Ceramic insulation Set of 6 elements held in place with support brackets Painted back panel Front insulation plug SiC target cavity Control thermocouple Stainless steel front plate Fig 3 Assemb...

Страница 15: ...tions for the LANDCAL Blackbody Type R1500T 220 240V CA970250 Fig 5 Element connections for the LANDCAL Blackbody Type R1500T 110 120V CA970251 Term inal Post Element Braid Element Head Element Suppor...

Страница 16: ...rocedure 2 When this check is completed two displays will be visible Upper display Actual source temperature Lower display Set point temperature 3 To modify the set point press the Up or Down button u...

Страница 17: ...n or 25 pin to connect to the computer RS232 Cable Furnace to PC connections Furnace end of cable female Computer end of cable male 9 pin 9 pin 25 pin Rx 3 3 2 Tx Tx 2 2 3 Rx Com5 5 7 Com 7 8 4 5 Link...

Страница 18: ...The controller s digital indicator on the side of the furnace displays the temperature measured Agreement between the thermocouple measurement and a reference standard radiation thermometer is better...

Страница 19: ...dard radiation thermometer 7 2 Thermometer calibration When the source is stable at the chosen control temperature place the thermometer under test in the holder and measure the thermometer output Not...

Страница 20: ...elements must be changed In the event of an element failure replace it in the manner described in Section 5 2 It is advisable to replace all elements if only one or two fail because new elements have...

Страница 21: ...ed with a heatsink pad All remnants of the old heatsink paste must be removed The thin white heatsink pad must be fitted between the solid state relay and the plate to which it is mounted 5 Tighten th...

Страница 22: ...e of operation These hidden features contain factory set parameters which should not be changed The Up and Down keys are used to alter the setpoint temperature in Level 1 operation and parameter value...

Страница 23: ...ramp rate allowing heating and cooling at the maximum rate OFF is the default setting 9 4 Altering the Power Limit when applicable 1 Press the Scroll key until OP Hi Output High is displayed 2 Press...

Страница 24: ...res The following spare parts are available for use with the Landcal Type R1500T Description Land Part N Silicon carbide target 135 161 Set of 6 heating elements 220 240 Volts 135 160 Set of 6 heating...

Страница 25: ...ing case the bench to mid target dimension equals that of the optical bench assembly Note 2 Mounted onto the 36in 915mm long optical bench are vertical and horizontal adjustment positioners which allo...

Страница 26: ...Landcal Blackbody Source Type R1500T User Guide Page 20...

Страница 27: ...ters accessories and mountings and special instruments based on UNO Application dedicated processors based on LANDMARK Graphic ABTS S and ABTS U FTS VDT S and VDT U DTT FLT5 A This 36 month warranty i...

Страница 28: ...onditions outside the Ambient Temperature Humidity and Vibration rating of the product environmental contamination solvent vapours deposition of airborne contamination cooling liquids of non neutral p...

Отзывы: