1  Introduction

This Installation Guide provides information on how to remove and replace the 

Main PCB (Part Nº 804320) in a Land Instruments International Model 4500 MkIII 

Continuous Opacity Monitoring System (COMS).

CAUTION

This procedure is intended for skilled instrument technicians and 

should be carried out in a clean workshop.
Take appropriate precautions to prevent damage from electrostatic 

discharge.

2  Removing the transceiver cover

1)  Using a 3mm Allen key, remove 

the 8 screws that secure the 

instrument cover.

2)  Slowly pull back the red cover 

until the keyboard connector 

can be seen. Disconnect the 

keyboard ribbon, and then 

remove the cover.

Содержание LAND 4500 MkIII

Страница 1: ...Land Instruments International 2010 2017 Installation Guide Model 4500 MkIII Main PCB Part N 804320 Publication N 809352 Language English Issue 2 28 July 2017...

Страница 2: ...e user documentation supplied before installing and operating the equipment The safety of any system incorporating this equipment is the responsibility of the assembler Protective Clothing Face and Ey...

Страница 3: ...ent damage during transit You must include a written report of the problem together with your own name and contact information address telephone number email address etc Design and Manufacturing Stand...

Страница 4: ...Model 4500 MkIII Main PCB Installation Guide...

Страница 5: ...Removing the transceiver cover 1 3 Replace the circuit board 2 4 Adjust the LED intensity 3 4 1 Adjusting the LED with a Modbus Connection 3 4 2 Adjusting the LED without a Modbus Connection 4 5 Repla...

Страница 6: ...Model 4500 MkIII Main PCB Installation Guide...

Страница 7: ...ng System COMS CAUTION This procedure is intended for skilled instrument technicians and should be carried out in a clean workshop Take appropriate precautions to prevent damage from electrostatic dis...

Страница 8: ...ke sure you can identify each of the connectors before you disconnect them If in doubt label each connector 2 Using a 3 mm Allen key remove the 6 mounting screws and remove the old circuit board 3 Fit...

Страница 9: ...Connector C 2 Connect a Modbus master to the RS485 port green yellow and black wires 3 Apply 24 V dc power to the red and blue wires of Connector A 4 Observe Modbus registers 121 122 and 124 These in...

Страница 10: ...instrument cover taking care not to short circuit any components inside the Model 4500 MkIII 6 Adjust potentiometer PR1 until Register 124 has a maximum value not exceeding 56000 7 Remove 24 V dc pow...

Страница 11: ...ite cord seal is correctly seated before pushing the cover home 3 Secure the cover with the eight screws 4 Apply 24 V dc power and check display and keyboard functionality 6 Reconfigure and Recalibrat...

Страница 12: ...erials or workmanship but due to application conditions Land Instruments International reserve the right to charge additionally for all labour on establishing this at the customer s works Warranty cov...

Страница 13: ......

Отзывы: