6.0 Technical 

specifications 

Sensor specifications: 
 
Sensor type 

:  Platinum sensor Pt100.  

 

 

α

 = 0.00385 

Sensor length 

:  150 mm 

Temperature range  :  -50°C to 400°C (-58°F to  
  

752°F) 

Accuracy : 

 Repeatability  : 

 0.005°C 

 Hysteresis

1)

 

:  0.01°C @ 0°C 

 Stability

2)

 

:  typ. 0.014°C @ 0°C 

Self heating effect  :  0.06°C/mW 

Diameter : 

OD3 

mm 

Immersion depth 

:  40 mm 

Media compatibility  :  INCONEL 600 
 

1)When used in the range –45°C to 400°C (-49°F to 752°F)

 

2)Stability when exposed to 400°C (752°F) for 100 hours.  
Stability will depend on actual use of the sensor. 

Response time 

:  A: 

τ

(50%)

 

= 5 sec. 

 

 

 

 

τ

(90%)

 

= 15 sec. 

Содержание JOFRA STS-103 B

Страница 1: ...User manual JOFRA STS 103 B Probe 150 Copyright 2006 AMETEK DENMARK A S...

Страница 2: ......

Страница 3: ...ontents Page 1 0 General information 4 2 0 Introduction 5 3 0 Functionality 6 3 1 Functional description 6 3 2 Connections 6 3 3 Serial number 8 4 0 Operation 9 4 1 Operation area 9 5 0 Maintenance 10...

Страница 4: ...General information This manual is only effective for the following product JOFRA STS 103 B 150 mm The product is manufactured by AMETEK Denmark A S GYDEVANG 32 34 DK 3450 ALLER D DENMARK TEL 45 48 1...

Страница 5: ...calibration and temperature measurement with AMETEK DTI and ATC systems and is ready for use Please read this manual carefully before use to obtain maximum value of your calibration system Warning Rea...

Страница 6: ...52 F Sensors may be supplied with certificates for a limited temperature range The resistance of the JOFRA STS 103 B sensor is converted to temperature according to IEC 751 ITS 90 calculated coefficie...

Страница 7: ...125660 00 26 01 2006 7 Model with banana plugs 2 meter cable Use the ground terminal in order to reduce noise Model with LEMO connection 2 meter or 0 5 meter cable R R B B G...

Страница 8: ...8 26 01 2006 125660 00 3 3 Serial number The serial number is placed on the sensor as shown on the figure below Serial number...

Страница 9: ...0 Operation 4 1 Operation area The sensor is intended for use in areas which meet the following Ambient temperature range 20 C to 70 C 4 F to 158 F Humidity 0 to 90 Protection class IP 50 Warning Do n...

Страница 10: ...user may carry out the following procedure himself Cleaning sensor Use alcohol or water and a soft cloth Caution The sensor must always be protected against any mechanical damage The sensor must never...

Страница 11: ...752 F Accuracy Repeatability 0 005 C Hysteresis1 0 01 C 0 C Stability2 typ 0 014 C 0 C Self heating effect 0 06 C mW Diameter OD3 mm Immersion depth 40 mm Media compatibility INCONEL 600 1 When used i...

Страница 12: ...icate The sensor is supplied with an accredited certificate according to the ITS 90 temperature scale The sensor is as standard calibrated in the range 45 C to 400 C 49 F to 752 F Calibration is carri...

Отзывы: