A sanitary sensor is normally similar to one of the two examples
shown below.
Fig. 1
Due to the very short insertion length, heat transfer through the
sensor tube and flange is of great importance to the calibration result.
Therefore, it is very important to achieve a good thermal contact to
the sensor as well as to the flange.
Make sure that the sensor fits exactly into the supplied insertion tube.
It is very important to have good mechanical and thereby also thermal
contact between the flange and the horizontal surface of the insertion
tube. In other words, there must be no air between the flange and the
insertion tube. (See also fig. 2).
8.2
The reference sensor
Included in the package is a STS-102 A 030 reference probe
especially developed for the purpose. It is important that this probe is
treated careful so it does not suffer any harm. When placing the
probe in the reference hole of the insert (Ø 4.2 mm) make sure that
there is no mechanical resistance.