7.2.2.5 Adjustment CJC ohm range
1. Connect the reference resistor to terminal 4 (+) and 1 (-).
2. Send command “
FUNC CJC_OHMS
”.
3. Send command “
NUM_CALP 2
”.
4. Send command “
REF[0] 900.000
”. (Replace 900.000 with actual
resistor value.)
5. Send command “
REF[1] 1100.000
”. (Replace 1100.000 with
actual resistor value.)
6. Connect 900 ohm reference resistor to DTI050.
7. Send command “
CAL_POINT[0]
”.
8. Connect 1100 ohm reference resistor to DTI050.
9. Send command “
CAL_POINT[1]
”.
7.2.2.6 Adjustment 500000 ohm range
1. Connect the reference resistor to terminal 3 (S+), 4 (I+), 2 (S-)
and 1 (I-).
2. Send command “
FUNC HI_OHMS
”.
3. Send command “
NUM_CALP 3
”.
4. Send command “
REF[0] 100.000
”. (Replace 100.000 with actual
resistor value.)
5. Send command “
REF[1] 300000.000
”. (Replace 300000.000 with
actual resistor value.)
6. Send command “
REF[2] 500000.000
”. (Replace 500000.000 with
actual resistor value.)
7. Connect 15 ohm reference resistor to DTI050.
8. Send command “
CAL_POINT[0]
”.
9. Connect 100 ohm reference resistor to DTI050.
10. Send command “
CAL_POINT[1]
”.
11. Connect 400 ohm reference resistor to DTI050.
12. Send command “
CAL_POINT[2]
”.