Switching off the calibrator
The following routine must be observed before the insertion tube is
removed and the instrument switched off (cf. fig. 1):
1.
If the calibrator has been heated up to temperatures above
50°C/122°F, you must wait until the instrument reaches a
temperature
below 50°C/122°F
before you switch it off.
2.
If the calibrator has reached a temperature below 0°C/32°F,
it should be heated momentarily to a temperature of
100°C/212°F.
3.
Switch off the calibrator using the power control switch (pos.
7).
4.
Remove the insertion tube from the calibrator using the tool
supplied with the instrument.
5.
Optional:
Store the calibrator in the carrying case.