Troubleshooting
23
4.2
Testing Electronic Unit
This test is only a test of the electronic unit for troubleshooting
purposes, and does not serve as a Verify or Certify test of the
complete system.
Use the following steps to test the electronic unit:
1. Be sure the environment is safe before removing the lid from
the housing.
2. If possible to access the sensing element with the material
below the sensor, or remove the IntelliPoint from the vessel,
use your finger to touch TP1 (Shown in Figure 2-4) while
holding any bare metal portion of the instrument housing
with the other hand. The system should go to its high level
state.
3. Again with no material touching the sensing element, touch
the tip of the sensing element with your finger, while holding
any bare metal portion of the instrument housing with the
other hand. The system should go to its high level state.
4. If the IntelliPoint changes to the high level state while
touching test point TP 1, but not when touching the tip of
the sensor, in most cases, the interconnecting cable is faulty.
See Section 4.6: Testing Integral Cable, or Section 4.7
Testing Remote Cable.
5. If The IntelliPoint is stuck in one state:
A. Remove power.
B. Disconnect coax cable that joins sensing element to
electronic unit.
C. Apply power.
D. Repeat steps 3 and 4.
E. If The IntelliPoint changes state with sensing element
disconnected, in most cases, sensing element is faulty.
See Section 4.5: Testing Sensing Element.
6. If there was no action in any of steps 2, 3, or 4:
A. Remove and then reapply power.
B. Press ReCal Button (
Shown in Figure 2-4
).
C. Observe that green LED flashes for about 60 seconds.
D. Green LED should be lit after 60 seconds.
E. Touch test point (
Shown in Figure 2-4
) with your
finger.
F. Alarm & Loop Current should change state. If so, circuit
board is working properly.
G. Reinstall instrument and press ReCal Button.
7. If The IntelliPoint fails all of above tests, in most cases
instrument is faulty. Use a replacement Input/Output
Module (IOM) or circuit board to determine fault.
Consult
factory
.
Содержание IntelliPoint RF RNT Series
Страница 4: ...Contents ...
Страница 6: ...Section 1 ...
Страница 37: ...Troubleshooting 31 Section 5 ...
Страница 40: ...RNTXX IntelliPoint RF Series User s Manual 34 Section 6 Control Drawings 6 1 FM FMc Control Drawings ...
Страница 41: ...Control Drawings 35 6 1 FM FMc Control Drawings Continued ...
Страница 42: ...RNTXX IntelliPoint RF Series User s Manual 36 6 1 FM FMc Control Drawings Continued ...
Страница 43: ...Control Drawings 37 6 1 FM FMc Control Drawings Continued ...
Страница 44: ...RNTXX IntelliPoint RF Series User s Manual 38 6 1 FM FMc Control Drawings Continued ...
Страница 45: ...Control Drawings 39 6 1 FM FMc Control Drawings Continued ...
Страница 46: ...RNTXX IntelliPoint RF Series User s Manual 40 6 1 FM FMc Control Drawings Continued ...
Страница 51: ...Control Drawings 45 6 1 FM FMc Control Drawings Continued ...
Страница 52: ...RNTXX IntelliPoint RF Series User s Manual 46 6 1 FM FMc Control Drawings Continued ...
Страница 56: ...RNTXX IntelliPoint RF Series User s Manual 50 6 2 ATEX Control Drawings ...
Страница 57: ...Control Drawings 51 6 2 ATEX Control Drawings Continued ...
Страница 66: ...RNTXX IntelliPoint RF Series User s Manual 60 6 4 IECEx Control Drawings ...
Страница 67: ...Control Drawings 61 6 4 IECEx Control Drawings Continued ...
Страница 75: ...Control Drawings 69 6 5 Mounting and Wiring for Spark Protector Drawings ...
Страница 76: ...RNTXX IntelliPoint RF Series User s Manual 70 6 5 Mounting and Wiring for Spark Protector Continued ...
Страница 77: ...Control Drawings 71 6 5 Mounting and Wiring for Spark Protector Continued ...
Страница 78: ...RNTXX IntelliPoint RF Series User s Manual 72 6 6 Adding a Padded Capacitor ...
Страница 79: ...Control Drawings 73 6 6 Adding a Padded Capacitor Continued ...
Страница 80: ...RNTXX IntelliPoint RF Series User s Manual 74 6 6 Adding a Padded Capacitor Continued ...
Страница 81: ...Control Drawings 75 6 7 Dual Seal Assembly for 700 Series Sensing Elements ...
Страница 82: ...RNTXX IntelliPoint RF Series User s Manual 76 6 7 Dual Seal Assembly Continued ...
Страница 83: ...Control Drawings 77 6 7 Dual Seal Assembly Continued ...
Страница 84: ...Appendix A ...