Installation, Operation and Maintenance

Optical Oxygen Analyzer

The 

OXYvisor

 is designed for use in manufacturing, industrial and laboratory analytical applications.  The installer should consult national, local 

codes and any site specific installation requirements to ensure governing regulations are met.  The 

OXYvisor

 goes through functional, performance 

and quality testing prior to leaving the factory.  For safe operation, prior to installation, read the entire manual and adhere strictly to all the 

WARNINGS

CAUTIONS

, and 

NOTES

 contained within this manual. Failure to do so could result in serious injury to the operator, the equipment or property.

1.4 Safety Summary

If the equipment is used in a manner not specified, the protection provided by the equipment may be impaired!

Si l’équipement est utilisé d’une manière non spécifiée, la protection fournie par l’équipement peut être altérée!

!

WARNING

!

WARNING

The 

OXYvisor

 product contains materials that can be reclaimed and recycled.  Waste Electrical and Electronic Equipment (WEEE) should never be 

disposed of in a municipal waste system (residential trash).  The Wheelie Bin marking on this product is a reminder to dispose of the product properly 
after it has completed its useful life and been removed from service in order to recycle the metals, plastics, and other components.  

According the directive 2012/19/EU, the monitoring and control instruments marked with the WEEE symbol and reaching their end-of-life must not be 
disposed of with other waste.  The user must dispose of the WEEE to a designated collection point for recycling of WEEE or send them back to the local 
Barben Analytical authorized distributor or the designated PCS (producer compliance scheme).  

Sales direct to end-user in European Union Member States:

The responsibility lies within the importer of the goods into each Member State, in this case it becomes the end-user’s responsibility to dispose of the 
goods according to the regulations of that Member State.  Barben Analytical has no obligation to dispose of these goods.

Sales made via our authorized distributors in European Union Member States:

In the case where our local distributor, located in Member State, imports and takes ownership of the goods they will bear the responsibility for disposal of 
the goods according to the regulations of the own Member State.  Barben Analytical has no obligation to dispose of these goods.

1.5 Environmental Information (WEEE)

Jusqu’à 240 VAC peuvent être présents dans les boîtiers de l’analyseur. Arrêtez toujours la ou les sources d’alimentation avant 

d’effectuer une maintenance ou un dépannage. Seul un électricien qualifié doit effectuer les raccordements électriques et les 

vérifications à la terre.

Verify ground connections and continuity of all equipment before applying power.

Vérifiez les connexions à la terre et la continuité de tous les équipements avant d’appliquer le courant.

Instrument grounding is mandatory.  Performance specifications and safety protection are void if instrument is operated from an improperly grounded 

power source.  

1.4.1 Electrical Safety

1�4�2 Grounding

Up to 240 VAC may be present in the analyzer housings.  Always shut down power source(s) before maintenance or troubleshooting.  

Only a qualified electrician should make electrical connections and ground checks.

When installing electronic measurement or control equipment, general installation precautions should be observed:

• 

Select a site that is free from direct sunlight, extreme temperatures, or abrupt temperature variations.

• 

Select a site where the ambient air is free from corrosive gases or abrasive materials.

The equipment should not contact surfaces or enclosures subjected to severe vibration or conductive heat.  Protective mounts should be installed to 
isolate the equipment from excessive vibration and thermal conduction.

• 

Do not install analyzer near equipment emitting electromagnetic interference (e.g. AC pumps, motors, etc).

• 

Electrical wiring should be installed according to the National Electrical Code, local regulations and codes along with and in addition to any other 
applicable industry codes and regulations.

• 

The supply voltage should strictly adhere to the instrument specifications, be supplied from a stable reliable source, and be provided with proper 

ground connection(s).

• 

Signal connections should be made using shielded wiring.

• 

Signal, control, and interface wiring should be located separately from power supply lines.

1.4.3 General Installation Precautions

Section 1

General Info & Safety

Содержание BARBEN ANALYTICAL OXY visor

Страница 1: ...IOM Safety Installation Operation and Maintenance Manual Abridged Safety related part Optical Oxygen Analyzer BOA For use with BOSx Optical Sensors Optical O2 Products BOA ATEX IEC NEC CEC...

Страница 2: ...se of this analyzer However Barben Analytical assumes no responsibility for any inaccuracies that may be contained in this manual In no event will Barben Analytical be liable for direct indirect incid...

Страница 3: ...er 13 3 6 1 Protective Ground Earthing Screw 13 3 6 2 AC Powered Analyzer 13 3 6 3 DC Powered Analyzer 14 3 7 Wiring Input Outputs 14 3 8 Electrical Entries 15 3 9 Securing the Rear Wiring Compartment...

Страница 4: ...ne 1 HazLoc Installation Control DWG 2P0345 p 2 16 Figure 13 3 Zone 1 HazLoc Installation Control DWG 2P0345 p 3 17 Figure 14 1 Zone 2 HazLoc Installation Control DWG 2P0346 p 1 17 Figure 14 2 Zone 2...

Страница 5: ...r a factory qualified local representative see back page for complete contact information Section 1 General Information Safety 1 1 Introduction Purpose of this Manual 1 2 About the OXYvisor Product Ov...

Страница 6: ...al distributor located in Member State imports and takes ownership of the goods they will bear the responsibility for disposal of the goods according to the regulations of the own Member State Barben...

Страница 7: ...galvanic isolation 4 to 20 mA Active Linear or Bi Linear 24 Vdc Digital Outputs Alarm Relays 4 Programmable relays optically isolated passive 24Vdc 0 05A pilot duty 0 45 A general use resistive load...

Страница 8: ...r Overall Dimensions 11 4 288mm DISPLAY AND THROUGH THE GLASS TOUCH KEYPAD 3 4 FNPT X2 or M25 x 1 5 6H conduit entries 1 2 FNPT X1 11 7 298mm FIBER OPTIC BUSHING P N B3907 1SFP or B3907 2SFP JAM NUT F...

Страница 9: ...e 1 The product label identifies important information and attributes of the specific model and should be reviewed in detail prior to installation Class I Division 2 Groups A B C D T4A US NEC Standard...

Страница 10: ...594 0 689 RoHS Material 0 020 0 025 thick 316 Stainless Steel Hole Diameter 0 150 4 typ Corners R 0 125 4 typ Model Serial Power BOA DC 1 SFP B SI YYMXXXX 21 6 26 4 VDC 5 Watts 1 Product and Order Spe...

Страница 11: ...61010 1 July 31 st 2019 Carson City NV USA Date and Place of Issue Melvin W Thweatt Jr Product Line Manager 1906 AME BA OV DoC EU R1 Potentially Explosive Potentiellement Explosif Zone 2 Model Select...

Страница 12: ...Suggested tools to complete the installation are listed below M8 or Small crescent Wrench Used for 2 pipe mount bracket installation Large flathead Screwdriver Used to mount analyzer to mounting brack...

Страница 13: ...e analyzer enclosure must be connected Requirements for the Protective Ground conductor are as follows The protective conductor must be of equal or greater size than any other current carrying wiring...

Страница 14: ...ve PT100 PT1000 PT1000 Temperature Sensor PT1000 positive wire then jumper to PT100 terminal below PT100 V PT1000 4 wire 3 rd wire or jumper to above terminal PT100 4 wire RTD input Drive Voltage posi...

Страница 15: ...pose Class I Div 2 or Zone II 3G and BOSx SENSOR WITH INTEGRAL RTD ANALYZER RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR WITH INTEGRAL RTD SEE BOSx...

Страница 16: ...NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JUNCTION BOX MANUFACTURER IS LIMATHERM S A CONTROL DRAWING 2 Z L2453 COATING SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER THE BOTTOM JUNCTION BOX PORT IS...

Страница 17: ...Control DWG 2P0346 p 1 1 OF 3 C SHIELDS OXYvisor HAZARDOUS LOCATION INSTALLATION DRAWING ZONE 2 UNLESS OTHERWISE SPECIFIED ALL 015 FRACTIONAL 000 005 00 010 0 015 APVD BY DRAWN BY DRAWING NO DATE SCA...

Страница 18: ...LE SIZE SHEET PART NO MATERIAL DESCRIPTION APVD DWN DESCRIPTION DATE REV REVISIONS ANGLES 1 4 TOLERANCES DIMENSIONS IN INCHES ECO 5200 Convair Drive Carson City NV 89706 PH 775 883 2500 FAX 775 297 47...

Страница 19: ...4 M5 SPU CW OR XD DR120WIN 4 M25 M25 M25 M25 M24 M5 SPU CW ATEX CERTIFICATION NO FTZ 08 ATEX 0182U ATEX MARKING II 2G Ex db IIC Gb IEC Ex CERTIFICATION NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JU...

Страница 20: ...S I DIVISION 2 HAZARDOUS AREA 1 Uo 2 5 V Vmax Uo Io 0 5 mA Imax Io Co 4 92 nF Ci Ccable Co Lo 16 40 H Li Lcable Lo USER SUPPLIED WIRING AND COMPONENTS 1 WIRING TO BE INSTALLED PER LOCAL CODE WIRING MU...

Страница 21: ...unauthorized operation was performed over Modbus Check the Modbus master device If a Modbus to USB cable is being used check cable requirements and replace N36 Unauthorized register was accessed via...

Страница 22: ...gone a reset undergone a reset Power OFF and ON If error is still present consult the factory Power OFF and ON If error is still present consult the factory X34 X34 This error indicates that some comp...

Страница 23: ...yzer Danger of explosion in areas subject to explosion hazard Connect the device correctly after maintenance Close and seal the device after maintenance work Flameproof joints are not intended to be r...

Страница 24: ...s nearly 14 000 colleagues at over 120 manufacturing locations around the world Supporting those operations are more than 80 sales and service locations across the United States and in more than 30 ot...

Отзывы: