Landcal Blackbody Source
Type P1600B2
User Guide
Page 4
3 Specification
Maximum working temperature:
1600°C / 2900°F
Recommended temperature range:
500 to 1550°C / 950 to 2825°F
Heating rate:
1.5 hours to 1400°C / 2550°F
Stability:
With the source controlling at temperature, the radiance
temperature will vary by less than 0.5°C over a 60 minute
period
Uniformity:
The temperature gradients across the middle 40mm of the
50mm cavity are within ±1°C over the temperature range 500
to 1600°C
Radiation cavity
Material:
Silicon Carbide
Inner diameter:
50mm / 2.0in approx.
Internal length:
300mm / 12.0in
Sighting extension tube: Length:
100mm / 4.0in
Inner diameter:
49mm / 1.9in
Emissivity:
0.998
Resistance heating element(s):
6 off SiC 151/356/20/25.4/7.4
Part Nº 135.017
Control thermocouple type:
Pt 13% Rh/Pt (Type R). 280mm long. Wire diameter 0.35mm.
(Master and overtemperature)
Twin bore insulation with end 60mm insulated with fish
spinebeads. This fits into an Alumina sheath of 6mm internal
diameter, 10mm outer diameter and length of 225mm. If
the thermocouple is replaced, the sheath should also be
replaced.
Controller
Master: Eurotherm 3216 with RS 232 serial interface
Over temperature: Eurotherm 2132
Electrical supply:
Single phase, 220/240V a.c. 50 to 60 Hz
Part No. 800358
Three phase, 4 wire, 415V a.c.
Part No. 800359
Three phase, 4 wire, 380V a.c.
Part No. 800360
Measuring thermocouple (if supplied) Type:
B(6/30) or R(0/13) or S(0/10)
Length:
600mm / 23.6in
Sheath:
8 x 5 x 600mm
Overall dimensions
Base length:
570mm / 22.4in
Base width:
350mm / 13.8in
Base height:
220mm / 8.7in
Heater enclosure diameter: 500mm / 19.7in
Heater enclosure length: 860mm / 33.8in
Bench to tube centre height:
470mm / 18.5in
Nett weight:
62kg / 135lb
The input a.c. power supply required for the particular furnace is specified on the furnace identity plate.
Mains supply requirements:
For single phase operation, 47 amps º 11.3kW
For 3 phase operation, 16 amps per phase = 3.8 kW per phase
Power consumption:
Furnace controller fitted with power turndown so it will not
draw maximum current available. For single phase operation
maximum consumption will be 7.0kW. For 3 phase operation
maximum consumption will be 2.3kW per phase.