Amersham Pharmacia Biotech INV-907 Скачать руководство пользователя страница 1

1 Introduction

Valve INV-907 is a motorised rotary 7–port
valve. The valve is used in ÄKTA

design

chromatography systems. It is powered from
the ÄKTAdesign P-900 series system pump, and
is controlled from UNICORN

control system.

The valve has 3 positions used for:
- loading a sample loop
- injecting the sample onto the column
- washing the system pump.

The valve can also be used to switch between
upflow and downflow in a column.

Features:

Completely swept flow path minimises
eluent or sample ”memory effect”.

Flow rates up to 100 ml/min.

All wetted parts are plastic PEEK.

Safety

2 Installation

Unpacking

Unpack the valve and check the items against
the packing list. Inspect the items for obvious
damage which may have occurred during
transportation. 

Installing the valve

1 Connect the valve with

two UniNet cables as a
part of the UniNet 2
chain.

2 Set the ID-switch at the

bottom to

the required valve number 

0–9

. The

number should correspond to that used in
UNICORN. 
Note: All valves must have different
numbers.

3 Mount the  valve vertically, in the way it

is shown in the user documentation of
your ÄKTAdesign chromatography
system.

4 Apply an identification number from the

labels supplied. The number should be the
same as that of the rotary switch at the
bottom of the valve. Place the number so
that it is easily readable from the front. 

5 Connect the tubing. 

CAUTION! Before connecting Valve
INV-907 ensure the power is switched OFF
at the system pump or the complete system.

WARNING! When using hazardous
chemicals, all suitable protective measures,
such as protective glasses, must be taken.

INV-907

3

Instructions

Valve INV-907

56-8101-01

Edition AC

UniNet 2 network

56-8101-01_AC.qxd  1998-10-26 11:45  Sida 1 (1,1)

Содержание INV-907

Страница 1: ...the valve 1 Connect the valve with two UniNet cables as a part of the UniNet 2 chain 2 Set the ID switch at the bottom to the required valve number 0 9 The number should correspond to that used in UN...

Страница 2: ...3 Operation The valve is controlled from UNICORN When used as an injection valve use the instruction InjectionValve in System Control Manual Flowpath The valve can be set in 3 positions LOAD position...

Страница 3: ...op volume Complete filling 2 5 times the sample loop volume Partial Filling Partial filling is used when high recovery is required Partial filling allows the injected volume to be changed without chan...

Страница 4: ...pprox 5 times the sample loop volume should be used to flush the loop and ensure that all sample is injected onto the column Storage Overnight The valve can be left filled with a buffer Weekend or Lon...

Страница 5: ...socket 2 Check the ID switch on the valve The ID number should correspond to the number set in UNICORN 3 Check the UniNet cable and replace if required The valve is switching to wrong position The val...

Страница 6: ...ysical data Internal volume Pos 1 LOAD Port 1 7 9 l Port 2 3 5 l Port 4 6 8 l Pos 2 INJECT Port 1 2 7 l Port 3 5 6 l Port 6 7 9 l Pos 3 WASH Port 2 4 8 l Port 5 7 9 l Flow channel diameter 0 8 mm Valv...

Страница 7: ...1 18 1113 81 Superloop 50 ml 1 18 1113 82 Superloop 150 ml 1 18 1023 85 Cable UniNet 0 7 m 1 18 1109 74 Mounting bracket 1 18 1109 11 Teflon tubing i d 1 8 o d 3 16 3 m 18 1112 47 Tubing connector for...

Страница 8: ...Biotech SE 751 84 Uppsala Sweden Trademarks KTA and UNICORN are trademarks of Amersham Pharmacia Biotech Limited or its subsidiaries Amersham is a trademark of Nycomed Amersham plc Pharmacia and Drop...

Отзывы: