Hoefer DALT System
Casting Gradient Gels
Amersham Biosciences
21
Light solution should begin to flow through the feed tube and mixer towards the
caster. At this point, a small amount of light solution can enter the caster.
10. When the light solution level in the gradient maker falls to a level about 1 cm above
the level of the heavy solution, open the heavy chamber exit tube clamp (b).
11. Watch the gradient enter the caster.
12. When the caster is filled to within 4 cm from the top of the cassettes, or the gradient
maker is empty, whichever comes first, turn off the pump and close the feed tube
clamp (c). Stop the pump before air enters the feed tube.
13. Pull the gradient feed tube out of the balance chamber grommet. Place its end in a
waste container to collect excess polymerizing acrylamide.
As soon as the feed tube is removed, the dense blue displacing solution flows down
the connecting tube to the unit. It should completely fill the V-well and the sloped
trough at the bottom of the caster. If the V-well is not completely filled and the level
of gel in the casettes is more than 1 cm below the top of the cassettes, you may add
up to 50 ml more displacing solution to the balance chamber. The gradient is now in
hydrostatic equilibrium in the unit, ready to polymerize (Figure 13). Apply overlay
as soon as possible (page 22).
Figure 13.
The gradient in
hydrostatic equilibrium
with the feed tube
removed. Displacing
solution fills the sloped
trough at the bottom of the
caster.
14. Quickly reopen clamp c and restart the pump to empty the gradient maker
completely of any excess polymerizing acrylamide from the gradient maker.
Collect the excess in a waste container. Dispose of unpolymerized acrylamide
according to applicable safety guidelines.
15. Rinse the gradient maker well to prevent polymerization within the tubing lines.
Place the feed tube in a larger waste vessel or a sink drain. Pour about a liter of
water into each chamber of the gradient maker and open all clamps.
16. Start the pump to flush the system. Flush 2 liters more water through the gradient
maker and tubing.
Displacing solution
Water
c